Zimmer recalls NexGen Stemmed Tibial Component Size 7 due to higher revision rates when used with specific LPS Flex femoral components in UK joint registry data. Affected patients should contact Zimmer for guidance.
Zimmer Biomet is conducting a voluntary medical device recall related to the NexGen Stemmed Option Tibial Components due to the clinically and statistically significant higher overall revision rates when these tibial components are used with either the Legacy Posterior Stabilized (LPS) Flex or LPS Flex Gender Solutions Femoral (GSF) components as compared to other total knee arthroplasties in the United Kingdom National Joint Registry (UK NJR). Removing the NexGen Stemmed Option Tibial Component f rom inventory will prevent its future implantation with either the LPS Flex or LPS Flex GSF femoral components and mitigate the increased revision risk with these two specific combinations of tibial and femoral component.
Zimmer is recalling a specific knee implant component because it has higher revision rates when paired with certain other knee implants in UK data. This means the component could need to be replaced more often than expected, which is risky for patients.
The recall is based on UK joint registry data showing these components have higher revision rates when used with specific L:PS Flex femoral implants. Revision means the implant needs to be removed and replaced, which can cause pain and longer recovery.
Patients who received this specific knee implant component during surgery in the UK are most at risk. The recall applies to all unexpired lots of this component.
Check if your knee implant is labeled 'NexGen Option Stemmed Tibial Component Size 7' with item number 00-5986-057-01. The recall applies to all unexpired lots of this component.
Reach out to Zimmer directly for guidance on next steps after recalling your implant.
The UK National Joint Registry found higher revision rates when this tibial component is used with specific LPS Flex femoral components.
No, this is a recall for surgical implants that may need replacement. Patients should contact Zimmer for guidance.
Look for the label 'NexGen Option Stemmed Tibial Component Size 7' with item number 00-5986-057-01 on your implant.
We’ve drafted a message you can send Zimmer to request your refund or repair — edit it as you like.
Subject: Recall Z-0470-2023 — Zimmer Zimmer Hello, I own a Zimmer that is covered by recall Z-0470-2023 from Zimmer. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.