Cardinal Health recalls 32Fr JP-2292 channel drains due to performance issues like lack of radiopacity and broken tips. Affected units have specific lot numbers and UDI/DI codes.
Cardinal Health recently received multiple complaints regarding various performance issues with Cardinal Health Jackson-Pratt Channel Drains including lack of radiopacity in wound drains, dull trocars impacting product performance, lack of flexibility of wound drains, and broken wound tips.
Cardinal Health is recalling 32Fr Jackson-Pratt Channel Drains with model JP-2:292 due to performance issues including lack of radiopacity, dull trocars, and broken tips. These issues could lead to improper wound drainage and potential complications during medical procedures.
The drains may not show up clearly on X-rays (lack of radiopacity), have dull trocars that affect performance, and break easily, leading to improper wound drainage during medical procedures.
Medical professionals using these drains for wound drainage procedures are most at risk. Patients receiving these drains may experience improper drainage or complications.
Check for the model number JP-2292 on the drain, lot numbers N210543 through N220163, and UDI/DI codes 50885380182684(cs) or 20885380182683(ea).
Identify the model number (JP-2292), lot numbers (N210543 to N220163), and UDI/DI codes (50885380182684(cs) or 20885380182683(ea)) to confirm if your drain is affected.
Cardinal Health received complaints about performance issues including lack of radiopacity, dull trocars, lack of flexibility, and broken tips.
Check the product details to see if your drain is affected; if it is, contact Cardinal Health for further instructions.
The drains may not show up clearly on X-rays, have dull trocars that affect performance, and break easily, potentially leading to improper wound drainage.
We’ve drafted a message you can send Cardinal Health to request your refund or repair — edit it as you like.
Subject: Recall Z-0994-2023 — Cardinal Health Cardinal Health Hello, I own a Cardinal Health that is covered by recall Z-0994-2023 from Cardinal Health. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.