Skip to content
Use caution Medical device recall · FDA · Published January 20, 2023

Coltene Whaledent SciCan STATIS SL S10001 Dental Handpieces Recall: Loose Parts Risk

Coltene Whaledent recalls SciCan STATIS SL S10001 dental handpieces due to risk of loose turbine parts during use, which could be swallowed or inhaled. All units are affected.

🇺🇸 FDA recall #Z-1208-2023Medical deviceColtene Whaledent ›
Take action

Contact the company

Recalling firmColtene Whaledent Inc235 Ascot Pkwy · Cuyahoga Falls, OH 44223-3701
✎ GovInfoNow plain-language explainer

Coltene Whaledent is recalling SciCan STATIS SL S10001 dental handpieces because a screw connection may loosen during use, increasing the risk of swallowing or inhaling parts. This could cause serious injury if the parts are inhaled or swallowed. Read more at govinfonow.com/recall/59023-coltene-whaledent-scican-statis-sl-s10001-dental-handpieces-recall-loose.

Source Summary by GovInfoNow · govinfonow.com/recall/59023-coltene-whaledent-scican-statis-sl-s10001-dental-handpieces-recall-loose · Updated August 19, 2026 · FDA enforcement record: FDA #Z-1208-2023

Cite this page

“Coltene Whaledent SciCan STATIS SL S10001 Dental Handpieces Recall: Loose Parts Risk.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/59023-coltene-whaledent-scican-statis-sl-s10001-dental-handpieces-recall-loose. Based on FDA recall #Z-1208-2023 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1208-2023%22&limit=1).

Coltene Whaledent Inc is voluntarily recalling SciCan STATIS SL S10001 Dental Handpieces due to a screw connection hazard that may loosen and fall off during treatment, according to the FDA. The Food and Drug Administration announced the recall on Jan. 20, 2023.

How to tell if you have it

  • Brand: Coltene Whaledent
  • Product: SciCan STATIS SL S10001 Dental Handpieces
  • Model Number: S10001
  • UDI-DI: +ESANS100010
  • Serial Numbers: All Serial Numbers
  • Units affected: 3896 units

The hazard

The screw connection between the turbine head and the push button may loosen and fall off during treatment, increasing the risk of swallowing or inhaling parts of the turbine head or the push button. The FDA said the handpieces are intended for removing carious material, reducing hard tooth structure, cavity preparation, finishing tooth preparations, restorations, and polishing teeth.

Context

GovInfoNow's records show Coltene Whaledent Inc has five other recalls on record since April 13, 2018. Other recent recalls of similar products include the Canon Alphenix INFX-8000F X-Ray System Recall (Dec. 19, 2025), the Howmedica Osteonics MICS3 Angled Sagittal Saw Attachment Recall (Feb. 2, 2026), and the Canon Alphenix X-ray Systems Recall (Aug. 4, 2026).

How to tell if you have this one

Check for the model number S10001 on the handpiece. All units are affected, including those with UDI-DI +ESANS100010.

  • Brand
    Coltene Whaledent
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Check for Loose Parts

    Inspect the turbine head and push button connection for looseness before each use.

  2. Stop Using if Loose

    Immediately stop using the handpiece if parts feel loose or move unexpectedly.

Where it went and when

Where it was distributed

Distributed nationwide, and outside the U.S.

Alaska: yesAKMaine: yesMEVermont: yesVTNew Hampshire: yesNHWashington: yesWAIdaho: yesIDMontana: yesMTNorth Dakota: yesNDMinnesota: yesMNIllinois: yesILWisconsin: yesWIMichigan: yesMINew York: yesNYRhode Island: yesRIMassachusetts: yesMAOregon: yesORNevada: yesNVWyoming: yesWYSouth Dakota: yesSDIowa: yesIAIndiana: yesINOhio: yesOHPennsylvania: yesPANew Jersey: yesNJConnecticut: yesCTCalifornia: yesCAUtah: yesUTColorado: yesCONebraska: yesNEMissouri: yesMOKentucky: yesKYWest Virginia: yesWVVirginia: yesVAMaryland: yesMDDelaware: yesDEArizona: yesAZNew Mexico: yesNMKansas: yesKSArkansas: yesARTennessee: yesTNNorth Carolina: yesNCSouth Carolina: yesSCDistrict of Columbia: yesDCOklahoma: yesOKLouisiana: yesLAMississippi: yesMSAlabama: yesALGeorgia: yesGAHawaii: yesHITexas: yesTXFlorida: yesFL

FDA distribution: “Nationwide Foreign: Canada Germany Indonesia Ireland Lebanon Saudia Arabia Turkey” States read from FDA’s own wording; “not listed” means the notice does not name it.

Distributed nationwide, and outside the U.S.

Recall timeline

FDA published this recall 47 days after the firm started it.

1. Firm started the recall: Jan 20, 202312. FDA classified it Class II: Mar 2, 202323. Published in FDA’s Enforcement Report: Mar 8, 202332023
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-1208-2023 (January 20, 2023) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-1208-2023 to see it in context.
  • Plain-language summary written by qwen3:4b on August 19, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA