Intuitive Surgical, Inc. voluntarily recalled Patient Side Carts with instrument arms used in da Vinci surgical systems on Jan. 24, 2023, according to the FDA. The recall addresses potential insertion ballscrew failure that may cause uncontrolled carriage movement and tissue injury.
How to tell if you have it
- Brand: Intuitive Surgical
- Product: Patient Side Cart, REF: 380620, with instrument arms for da Vinci X Surgical System, REF: IS4200
- Product: Patient Side Cart, REF: 380652, with instrument arms for da Vinci Xi Surgical System, REF: IS4000
- Units affected: 9
The hazard
Preventative maintenance data identified instrument arms at risk of insertion ballscrew failure. This defect allows for uncontrolled carriage movement along the insertion axis, which may cause tissue injury.
Context
GovInfoNow's records show Intuitive Surgical has 53 other recalls on record since June 1, 2017. Recent similar recalls include the Da Vinci X and Xi Surgical Systems Recall regarding broken manipulator screws (April 2, 2026) and the Da Vinci 5 dV5 Tower NIR Handheld Camera Recall for unintentional activation (April 24, 2026). Other recent recalls of similar products include the IMRIS ORT300 Operating Room Table Recall for unintended movement risk (Aug. 6, 2026) and the Olympus HX-400U-30 Ligating Device Recall for unintentional loop fixation risk (Oct. 30, 2025).