The FDA announced on Feb. 7, 2023, that Stryker GmbH is recalling its Hoffmann LRF Transport Strut (Model Number 4933-0-400) due to a manufacturing defect. The hazard involves an oversized thread pitch that prevents proper locking of the distal ring fixation.
How to tell if you have it
- Brand: Stryker
- Product: Hoffmann LRF, Transport Strut
- Model Number: 4933-0-400
- Serial numbers: D32158, D32160, D32161, D32162, D32163, D32164, D34343, D34344, D40690, D41629, D41630, D41631, D44713, G16331, G16332, G16333, G26278, G29759, G29760, G32730, G32731, G32735, G45119, H30398, H30399, H66392
- Units affected: 488
The hazard
According to the FDA, the thread pitch is oversized at the distal end of the threaded rod. As a result, the distal ring fixation cannot be locked between 0-50mm of the threaded rod.
Context
GovInfoNow's records show Stryker GmbH has 20 other recalls on record since March 12, 2018. Other recent recalls of similar products include Stryker Hoffmann II Connecting Rod Micro 3 x 120mm (July 1, 2026), Stryker Gamma4 Right Long Nail Recall (May 28, 2025), and Stryker Femoral Nail Recall (May 22, 2025).