Skip to content
Lower risk Medical device recall · FDA · Published February 17, 2023

Raystation treatment planning system recall: Incorrect collimator angle in exported plans

Raysearch recalls Raystation radiation therapy software versions 9.0.0-13.1.0 with incorrect collimator angle in exported DICOM plans. Users may need to update software to prevent treatment errors.

🇺🇸 FDA recall #Z-1281-2023Medical deviceRaysearch ›
Take action

Contact the company

Recalling firmRAYSEARCH LABORATORIES ABEugeniavagen 18c · Stockholm, Sweden
✎ GovInfoNow plain-language explainer

Raysearch is recalling its Raystation radiation therapy software versions 9.0.0 through 13.1.0 because setting a non-zero collimator angle in the Virtual Simulation module can cause incorrect angles in exported treatment plans. This could lead to improper radiation delivery during therapy if not corrected. Read more at govinfonow.com/recall/58873-raystation-treatment-planning-system-recall-incorrect-collimator-angle-in-exported.

Source Summary by GovInfoNow · govinfonow.com/recall/58873-raystation-treatment-planning-system-recall-incorrect-collimator-angle-in-exported · Updated August 19, 2026 · FDA enforcement record: FDA #Z-1281-2023

Cite this page

“Raystation treatment planning system recall: Incorrect collimator angle in exported plans.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/58873-raystation-treatment-planning-system-recall-incorrect-collimator-angle-in-exported. Based on FDA recall #Z-1281-2023 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1281-2023%22&limit=1).

The FDA announced on Feb. 17, 2023, that Raysearch Laboratories AB is recalling Raystation treatment planning systems for radiation therapy due to a hazard involving incorrect collimator angles in exported plans. The agency said the recall affects 10,770 systems distributed nationwide.

How to tell if you have it

  • Brand: Raysearch
  • Product: Raystation treatment planning system
  • Model Numbers: 9.0.0, 9.0.1, 9.1.0, 9.2.0, 10.0.0, 10.0.1, 10.0.2, 10.1.0, 10.1.1, 11.0.0, 11.0.1, 11.0.3, 11.0.4, 12.0.0, 12.1.0, 12.1.1, 12.0.3, 12.1.2, 12.0.4, 12.1.3, 13.0.0, 13.1.0
  • Software Version Numbers: 9A, 9B, 10A, 10B, 11A, 11B and 12A including service packs
  • UDI-DI: 0735000201067920221007

The hazard

The FDA said it is possible to set a non-zero collimator angle in the Virtual Simulation module that appears correct in the user interface but is incorrect in the DICOM-exported RT plan. Specifically, the DICOM attribute Beam Limiting Device Angle (300A,0120) is set to zero regardless of the selected collimator angle when using the VSIM Export button.

Context

GovInfoNow's records show RAYSEARCH LABORATORIES AB has issued 36 other recalls since July 17, 2017. Recent brand recalls include a May 20, 2026 RayCare 2:024A SP1 Oncology Software recall for data loss risk and a Nov. 28, 2025 RayStation/RayPlan 10B Software Recall for radiation dose calculation risks. Other recent recalls of similar products include an Dec. 12, 2025 Elekta Leksell GammaPlan 11.1 Software Recall and a Feb. 6, 2026 Hitachi Proton Beam Therapy System Recall.

How to tell if you have this one

Check if your Raystation software version is listed in the model numbers 9.0.0 through 13.1.0 or software versions 9A through 12B. The recall includes specific UDI-DI numbers starting with 0735000201067920221007.

  • Brand
    Raysearch
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Check Software Version

    Verify your Raystation version by checking the software version number in the program interface or documentation.

  2. Update Software

    Install the latest software update from Raysearch to fix the collimator angle issue in exported plans.

Where it went and when

Where it was distributed

Distributed nationwide.

Alaska: yesAKMaine: yesMEVermont: yesVTNew Hampshire: yesNHWashington: yesWAIdaho: yesIDMontana: yesMTNorth Dakota: yesNDMinnesota: yesMNIllinois: yesILWisconsin: yesWIMichigan: yesMINew York: yesNYRhode Island: yesRIMassachusetts: yesMAOregon: yesORNevada: yesNVWyoming: yesWYSouth Dakota: yesSDIowa: yesIAIndiana: yesINOhio: yesOHPennsylvania: yesPANew Jersey: yesNJConnecticut: yesCTCalifornia: yesCAUtah: yesUTColorado: yesCONebraska: yesNEMissouri: yesMOKentucky: yesKYWest Virginia: yesWVVirginia: yesVAMaryland: yesMDDelaware: yesDEArizona: yesAZNew Mexico: yesNMKansas: yesKSArkansas: yesARTennessee: yesTNNorth Carolina: yesNCSouth Carolina: yesSCDistrict of Columbia: yesDCOklahoma: yesOKLouisiana: yesLAMississippi: yesMSAlabama: yesALGeorgia: yesGAHawaii: yesHITexas: yesTXFlorida: yesFL

FDA distribution: “US Nationwide” States read from FDA’s own wording; “not listed” means the notice does not name it.

Distributed nationwide.

Recall timeline

FDA published this recall 40 days after the firm started it.

1. Firm started the recall: Feb 17, 202312. FDA classified it Class II: Mar 22, 202323. Published in FDA’s Enforcement Report: Mar 29, 202332023
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report

Dates from the agency record only; steps without a date are left out.

✎ GovInfoNow

Questions people ask

What versions are affected?

Software versions 9.0.0 through 13.1.0 and software versions 9A through 12B, including service packs.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-1281-2023 (February 17, 2023) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-1281-2023 to see it in context.
  • Plain-language summary written by qwen3:4b on August 19, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA