Skip to content
Lower risk Medical device recall · FDA · Published February 28, 2023

Encore Medical EMPOWR Knee Drills: Incorrect Size Risk of Implant Loosening

Encore Medical recalls two EMPOWR knee drills due to incorrect diameter markings that could cause implant loosening or bone fractures during surgery.

🇺🇸 FDA recall #Z-1384-2023Medical deviceEncore Medical ›
Take action

Contact the company

Recalling firmEncore Medical, LP9800 Metric Blvd · Austin, TX 78758-5445
✎ GovInfoNow plain-language explainer

Encore Medical is recalling two specific knee drills because they have the wrong diameter markings. If surgeons use these incorrect drills during surgery, the implants might loosen, not fit properly, or cause broken bones. Read more at govinfonow.com/recall/58845-encore-medical-empowr-knee-drills-incorrect-size-risk-of-implant.

Source Summary by GovInfoNow · govinfonow.com/recall/58845-encore-medical-empowr-knee-drills-incorrect-size-risk-of-implant · Updated August 19, 2026 · FDA enforcement record: FDA #Z-1384-2023

Cite this page

“Encore Medical EMPOWR Knee Drills: Incorrect Size Risk of Implant Loosening.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/58845-encore-medical-empowr-knee-drills-incorrect-size-risk-of-implant. Based on FDA recall #Z-1384-2023 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1384-2023%22&limit=1).

The FDA announced on Feb. 28, 2023, that Encore Medical, LP is recalling EMPOWR Porous Knee Patella Over Drills and EMPOWR Knee Patella Drills due to incorrect diameter markings. The agency said the error poses a risk of implant loosening, improper fitting, or fractured bone during surgery.

How to tell if you have it

  • Brand: Encore Medical
  • Product: EMPOWR Porous Knee Patella Over Drill, REF: 802-05-104; Lot: 388620L01
  • Product: EMPOWR Knee Patella Drill, REF: 802-05-108; Lot: 388620L03

The hazard

The recall involves patella drills marked with an incorrect diameter size of 5.6mm instead of the required 5.4mm, and porous patella overdrills marked with 5.4mm instead of 5.6mm. According to the FDA, using the incorrect drill during surgery risks implant loosening, improper fitting, or fractured bone.

Context

GovInfoNow's records show Encore Medical, LP has issued 43 other recalls since May 31, 2017. Recent brand actions include a Feb. 16, 2026, product mix-up recall for EMPOWR 3D Knee Ins and a Jan. 15, 2026, labeling hazard recall for RSP Humeral Socket Inserts. Other recent recalls of similar products include U2 Total Knee System: Incorrect Implant Orientation Risk (March 19, 2026), Think Surgical TMINI Tracking Arrays: Risk of Misalignment Due to Loose Knob (May 8, 2026), and U2 Total Knee System: Risk of Improper Assembly Due to Manufacturing Defect (May 20, 2026).

How to tell if you have this one

Check for the model numbers 802-05-104 or 802-05-108 on the drill. The recall includes specific serial numbers and lot codes listed in the product description.

  • Brand
    Encore Medical
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Check Drill Model

    Verify the drill model number is 802-05-104 or 802-05-108 to confirm if it's recalled.

Where it went and when

Where it was distributed

Distributed in 8 states.

Alaska: not listedAKMaine: not listedMEVermont: not listedVTNew Hampshire: not listedNHWashington: yesWAIdaho: not listedIDMontana: not listedMTNorth Dakota: not listedNDMinnesota: not listedMNIllinois: not listedILWisconsin: not listedWIMichigan: not listedMINew York: not listedNYRhode Island: not listedRIMassachusetts: not listedMAOregon: not listedORNevada: not listedNVWyoming: not listedWYSouth Dakota: not listedSDIowa: not listedIAIndiana: yesINOhio: not listedOHPennsylvania: not listedPANew Jersey: not listedNJConnecticut: not listedCTCalifornia: yesCAUtah: not listedUTColorado: not listedCONebraska: not listedNEMissouri: not listedMOKentucky: not listedKYWest Virginia: not listedWVVirginia: not listedVAMaryland: yesMDDelaware: not listedDEArizona: not listedAZNew Mexico: not listedNMKansas: yesKSArkansas: not listedARTennessee: not listedTNNorth Carolina: not listedNCSouth Carolina: not listedSCDistrict of Columbia: not listedDCOklahoma: not listedOKLouisiana: not listedLAMississippi: not listedMSAlabama: yesALGeorgia: not listedGAHawaii: yesHITexas: yesTXFlorida: not listedFL

FDA distribution: “US: TX, IN, MD, CA, KS, AL, HI, WA” States read from FDA’s own wording; “not listed” means the notice does not name it.

States: Alabama, California, Hawaii, Indiana, Kansas, Maryland, Texas, Washington.

Recall timeline

FDA published this recall 50 days after the firm started it.

1. Firm started the recall: Feb 28, 202312. FDA classified it Class II: Apr 12, 202323. Published in FDA’s Enforcement Report: Apr 19, 202332023
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-1384-2023 (February 28, 2023) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-1384-2023 to see it in context.