Skip to content
Lower risk Medical device recall · FDA · Published May 4, 2023

Randox Immunoassay Premium Level 3: Delay in test results due to quality control issues

Randox recalls Immunoassay Premium Level 3 quality control kits for potential delays in test results due to high recovery values outside the acceptable range on Roche Cobas e801 analyzers.

🇺🇸 FDA recall #Z-2021-2023Medical deviceRandox ›
Take action

Contact the company

Recalling firmRandox Laboratories Ltd.55 Diamond Road · Crumlin (North), Ireland
✎ GovInfoNow plain-language explainer

Randox is recalling Immunoassay Premium Level 3 quality control kits that may cause delays in test results when used with Roche Cobas e801 analyzers. This happens because the test results show high recovery values outside the expected range, which can lead to inaccurate test reporting. Read more at govinfonow.com/recall/58459-randox-immunoassay-premium-level-3-delay-in-test-results-due.

Source Summary by GovInfoNow · govinfonow.com/recall/58459-randox-immunoassay-premium-level-3-delay-in-test-results-due · Updated August 19, 2026 · FDA enforcement record: FDA #Z-2021-2023

Cite this page

“Randox Immunoassay Premium Level 3: Delay in test results due to quality control issues.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/58459-randox-immunoassay-premium-level-3-delay-in-test-results-due. Based on FDA recall #Z-2021-2023 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2021-2023%22&limit=1).

The FDA announced on May 4, 2023, that Randox Laboratories Ltd. is recalling Immunoassay Premium Level 3 quality control material due to potential delays in reporting results. The hazard stems from quality controls running high outside the expected range for ACTH testing.

How to tell if you have it

  • Brand: Randox
  • Product: Immunoassay Premium Level 3-In vitro diagnostic use, as assayed quality control material to monitor the accuracy and
  • Catalog Number: IA2640
  • GTIN/UPC: 05055273203868
  • Lot Code: 2031EC
  • Expiration Date: 28th Nov 23
  • Quantity: 1 kit

The hazard

The recall addresses delays in reporting results because quality controls ran high outside of range. This issue affects the target value and range for ACTH in Immunoassay Premium Level 3, IA2640, lot 2031EC and Immunoassay Premium Control Tri Level kit batch 583135 lot 2031EC when used on the Roche Cobas e801 due to high recovery outside range.

Context

GovInfoNow's records show Randox Laboratories Ltd. has had 49 other recalls since June 7, 2017. Recent brand recalls include a June 7, 2024 recall for missing FDA 510(k) clearance and an April 5, 2024 recall for light leak risks in Evidence MultiSTAT Software v3.7. Other recent recalls of similar products include the Roche Elecsys Anti-TSHR Assay Recall (Nov. 21, 2025), the Roche cobas pro integrated solutions recall (Jan. 20, 2026), and the Immunotech Estrone RIA Kit Recall (Dec. 15, 2025).

How to tell if you have this one

Check for the catalog number IA2640 and lot number 2031EC on the kit. The expiration date is November 28, 2023. These kits were sold with Roche Cobas e801 analyzers.

  • Brand
    Randox
  • Product
    Immunoassay Premium Level 3-In vitro diagnostic use, as assayed quality control material to monitor the accuracy and
  • Category
    Medical device
FDA recall event · 2 products

All 2 products in this recall

FDA lists each product line separately under one recall event. Check your product description, lot or code numbers against each row. Every row shares the hazard and remedy on this page.

  • Classification
    Class III
  • Status
    Ongoing
  • Distribution
    CA, IL, VA, WA, WI including PR
Recall numberProduct, codes & lotsQuantity
Z-2021-2023Immunoassay Premium Level 3-In vitro diagnostic use, as assayed quality control material to monitor the accuracy and reproducibility of analytes. Catalog Number: IA2640; GTIN:05055273203868 Batch/Lot Number: 2031EC Exp. Date: 28th Nov 231 kit
Z-2022-2023Immunoassay Premium Control Tri Level-In vitro diagnostic use, as assayed quality control material to monitor the accuracy and reproducibility of analytes Catalog Number: IA2633; GTIN: 05055273203837 Batch/Lot Number: 583135 Exp. Date: 28th April 2330 kits
Based on the official FDA remedy

What to do — step by step

  1. Check Kit Details

    Verify the catalog number (IA2640) and lot number (2031EC) on the kit to confirm it is recalled.

Where it went and when

Where it was distributed

Distributed in 5 states and Puerto Rico.

Alaska: not listedAKMaine: not listedMEVermont: not listedVTNew Hampshire: not listedNHWashington: yesWAIdaho: not listedIDMontana: not listedMTNorth Dakota: not listedNDMinnesota: not listedMNIllinois: yesILWisconsin: yesWIMichigan: not listedMINew York: not listedNYRhode Island: not listedRIMassachusetts: not listedMAOregon: not listedORNevada: not listedNVWyoming: not listedWYSouth Dakota: not listedSDIowa: not listedIAIndiana: not listedINOhio: not listedOHPennsylvania: not listedPANew Jersey: not listedNJConnecticut: not listedCTCalifornia: yesCAUtah: not listedUTColorado: not listedCONebraska: not listedNEMissouri: not listedMOKentucky: not listedKYWest Virginia: not listedWVVirginia: yesVAMaryland: not listedMDDelaware: not listedDEArizona: not listedAZNew Mexico: not listedNMKansas: not listedKSArkansas: not listedARTennessee: not listedTNNorth Carolina: not listedNCSouth Carolina: not listedSCDistrict of Columbia: not listedDCOklahoma: not listedOKLouisiana: not listedLAMississippi: not listedMSAlabama: not listedALGeorgia: not listedGAHawaii: not listedHITexas: not listedTXFlorida: not listedFLPuerto Rico: yesPR

FDA distribution: “CA, IL, VA, WA, WI including PR” States read from FDA’s own wording; “not listed” means the notice does not name it.

States: California, Illinois, Puerto Rico, Virginia, Washington, Wisconsin.

Recall timeline

FDA published this recall 2 months after the firm started it.

1. Firm started the recall: May 4, 202312. FDA classified it Class III: Jun 28, 202323. Published in FDA’s Enforcement Report: Jul 5, 202332023
  1. Firm started the recall
  2. FDA classified it Class III
  3. Published in FDA’s Enforcement Report

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-2021-2023 (May 4, 2023) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-2021-2023 to see it in context.
  • Plain-language summary written by qwen3:4b on August 19, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA