The FDA announced on May 22, 2023, that CareFusion 303, Inc. is recalling BD Pyxis CII Safe automated dispensing cabinet devices due to unacceptable current leakage levels when plugged into multi-socket outlet power strips. The hazard poses a risk of system downtime and potential electric shock to users.
How to tell if you have it
- BD Pyxis CII Safe, Software: V7, REF: 107-96
- BD Pyxis CII Safe, Software: V7, REF: 107-97
- BD Pyxis CII Safe, Software: V7, REF: 107-97-201
- BD Pyxis CII Safe, Software: V7, REF: 107-223
- BD Pyxis CII Safe, Software: V8, REF: 111-199-01
- BD Pyxis CII Safe, Software: V8, REF: 111-200-01
- BD Pyxis CII Safe, Software: V8, REF: 111-201-01
- BD Pyxis CII Safe, Software: V8, REF: 111-202-01
- BD Pyxis CII Safe, Software: V8, REF: 111-203-01
- BD Pyxis CII Safe, Software: V8, REF: 111-204-01
- BD Pyxis CII Safe, Software: V8, REF: 111-205-01
- BD Pyxis CII Safe, Software: V8, REF: 111-206-01
- BD Pyxis CII Safe, Software: V8, REF: 107-245-01
- BD Pyxis CII Safe, Software: V8, REF: 107-246-01
- BD Pyxis CII Safe, Software: V9, REF: 107-252-01
- BD Pyxis CII Safe, Software: V9, REF: 111-215
- BD Pyxis CII Safe, Software: V9, REF: 111-216
- BD Pyxis CII Safe, Software: V9, REF: 111-219
- BD Pyxis CII Safe, Software: V9, REF: 111-221
- BD Pyxis CII Safe, Software: ES, REF: 107-254-01
- BD Pyxis CII Safe, Software: ES, REF: 107-255-01
- BD Pyxis CII Safe, Software: ES, REF: 1116-00
- BD Pyxis CII Safe, Software: ES, REF: 1117-00
- BD Pyxis CII Safe, Software: N/A, REF: 111-113
- BD Pyxis CII Safe, Software: N/A, REF: 111-114
- BD Pyxis CII Safe, Software: N/A, REF: 111-115
- BD Pyxis CII Safe, Software: N/A, REF: 111-160-201
- BD Pyxis CII Safe, Software: N/A, REF: 111-217
- BD Pyxis CII Safe, Software: N/A, REF: 111-218
- BD Pyxis CII Safe, Software: N/A, REF: 111-220
- BD Pyxis CII Safe, Software: N/A, REF: 111-222
- Tripp Lite SUPER6TEL Surge Suppressor, REF: CN 200-025
The hazard
Automated dispensing cabinet devices plugged into any multi-socket outlet power strips may present unacceptable levels of current leakage. This condition could expose users to system downtime and/or potential electric shock.
Context
GovInfoNow's records show CareFusion 303, Inc. has 64 other recalls on record since 2017-06-12. Other recent recalls of similar products include BD Pyxis MedStation ES Tower 352 Medication Cabinet Recall (2026-04-30), CareFusion - CareFusion labeling recall (2026-02-11), and BD Pyxis ES Enterprise Server Software Update Issue Recall (2025-12-03). Other recent recalls of similar products include Philips Allura Xper FD10C Medical System Recall (2025-12-15), Philips Allura Xper FD10F Medical System Recall (2025-12-15), and Spacelabs Ultraview SL Command Module Software Recall (2026-03-26).