Skip to content
Use caution Medical device recall · FDA · Published June 5, 2023

Siemens Cios Select VA 21 Mobile X-ray System Recall: Risk of Electric Shock from Liquid

Siemens recalls Cios Select VA 21 mobile X-ray systems due to potential electric shock from liquid ingress if plexiglass cover falls off. Affected units have specific serial numbers and were sold for medical use. Contact Siemens for details.

🇺🇸 FDA recall #Z-2240-2023Medical deviceSiemens ›
Take action

Contact the company

Recalling firmSiemens Medical Solutions USA, Inc40 Liberty Blvd · Malvern, PA 19355-1418
✎ GovInfoNow plain-language explainer

Siemens is recalling Cios Select VA 21 mobile X-ray systems that could have incorrect glue causing the plexiglass cover to fall off. This might let liquid enter the system, creating a risk of electric shock if the user touches live parts. The recall affects medical devices used in operating rooms and other clinical settings. Read more at govinfonow.com/recall/57900-siemens-cios-select-va-21-mobile-x-ray-system-recall.

Source Summary by GovInfoNow · govinfonow.com/recall/57900-siemens-cios-select-va-21-mobile-x-ray-system-recall · Updated August 19, 2026 · FDA enforcement record: FDA #Z-2240-2023

Cite this page

“Siemens Cios Select VA 21 Mobile X-ray System Recall: Risk of Electric Shock from Liquid.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/57900-siemens-cios-select-va-21-mobile-x-ray-system-recall. Based on FDA recall #Z-2240-2023 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2240-2023%22&limit=1).

The FDA announced on June 5, 2023, that Siemens Medical Solutions USA, Inc. is recalling the Cios Select VA 21 with FD-A Mobile X-ray system due to a risk of electric shock. The hazard stems from incorrect glue on the plexiglass sheet potentially allowing liquid ingress.

How to tell if you have it

  • Brand: Siemens
  • Product: Cios Select VA 21 with FD-A Mobile X-ray system
  • Material #: 11515088
  • UDI-DI: 04056869148243
  • Serial Numbers: 30001, 30002, 30008, 30013, 30029, 30031, 30032, 30033, 30034
  • Units affected: 9

The hazard

Incorrect glue was applied to the plexiglass sheet on the single tank cover assembly, which could fall off and allow liquid ingress. This creates a risk of electric shock if the user touches liquid connected to internal live parts while the plexiglass drops off.

Context

GovInfoNow's records show Siemens Medical Solutions USA, Inc. has 182 other recalls on record since June 6, 2017. Other recent recalls of similar products include GE Allia Moveo X-ray System Recall (April 20, 2026), GE Medical Systems Allia IGS 3 Pulse X-Ray System Recall (April 20, 2026), and GE Medical Systems Allia IGS Pulse X-ray System Recall (April 20, 2026).

How to tell if you have this one

Check for the serial numbers 30001, 30002, 30008, 30013, 30029, 30031, 30032, 30033, or 30034 on the device. The system is intended for use in operating rooms, trauma, endoscopy, and other medical settings.

  • Brand
    Siemens
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Check Device Serial Number

    Look for the serial numbers 30001, 30002, 30008, 30013, 30029, 30031, 30032, 30033, or 30034 on the device.

Where it went and when

Where it was distributed

Distributed nationwide and abroad; the notice names 6 states.

Alaska: nationwideAKMaine: nationwideMEVermont: nationwideVTNew Hampshire: nationwideNHWashington: nationwideWAIdaho: nationwideIDMontana: nationwideMTNorth Dakota: nationwideNDMinnesota: named in the noticeMNIllinois: nationwideILWisconsin: nationwideWIMichigan: nationwideMINew York: named in the noticeNYRhode Island: nationwideRIMassachusetts: nationwideMAOregon: nationwideORNevada: nationwideNVWyoming: nationwideWYSouth Dakota: nationwideSDIowa: nationwideIAIndiana: nationwideINOhio: nationwideOHPennsylvania: nationwidePANew Jersey: nationwideNJConnecticut: nationwideCTCalifornia: nationwideCAUtah: nationwideUTColorado: nationwideCONebraska: nationwideNEMissouri: nationwideMOKentucky: nationwideKYWest Virginia: nationwideWVVirginia: nationwideVAMaryland: nationwideMDDelaware: nationwideDEArizona: nationwideAZNew Mexico: nationwideNMKansas: named in the noticeKSArkansas: nationwideARTennessee: nationwideTNNorth Carolina: named in the noticeNCSouth Carolina: nationwideSCDistrict of Columbia: nationwideDCOklahoma: nationwideOKLouisiana: nationwideLAMississippi: nationwideMSAlabama: nationwideALGeorgia: named in the noticeGAHawaii: nationwideHITexas: nationwideTXFlorida: named in the noticeFL

FDA distribution: “Worldwide - US Nationwide distribution in the states of FL, GA, KS, MN, NC, NY and the countries of Austria, Bangladesh, Brazil, Canada, China, Croatia, El Salvador, France, Germany, Indonesia, Italy, Japan, Jordan, Lebanon, Luxembourg,…” States read from FDA’s own wording; “not listed” means the notice does not name it.

Named in the notice: Florida, Georgia, Kansas, Minnesota, New York, North Carolina.

Recall timeline

FDA published this recall 58 days after the firm started it.

1. Firm started the recall: Jun 5, 202312. FDA classified it Class II: Jul 27, 202323. Published in FDA’s Enforcement Report: Aug 2, 202332023
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-2240-2023 (June 5, 2023) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-2240-2023 to see it in context.
  • Plain-language summary written by qwen3:4b on August 19, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA