Lower risk Medical device recall · FDA · Published June 6, 2023

Murata Vios BSM2050 Bedside Monitor Software Version 1.54.20/1.56.00 Recall

Murata Vios recalls bedside monitors with specific software versions due to potential incorrect pulse oximeter warnings. Affected units may display false 'Pulse Ox Sensor Off' alerts.

🇺🇸 FDA recall #Z-1169-2025Medical deviceMurata Vios ›
Take action

Contact the company

Recalling firmMurata Vios, Inc.700 Commerce Dr Ste 190 · Woodbury, MN 55125-9239
✎ GovInfoNow plain-language explainer

Murata Vios is recalling bedside monitors with software versions 1.54.20 and 1.56.00 because a faulty sensor can cause false warnings about low oxygen levels. This could lead to unnecessary alarm or delayed medical response. Read more at govinfonow.com/recall/57887-murata-vios-bsm2050-bedside-monitor-software-version-1-54-20.

Source Summary by GovInfoNow · govinfonow.com/recall/57887-murata-vios-bsm2050-bedside-monitor-software-version-1-54-20 · Updated August 19, 2026 · FDA enforcement record: FDA #Z-1169-2025

Cite this page

“Murata Vios BSM2050 Bedside Monitor Software Version 1.54.20/1.56.00 Recall.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/57887-murata-vios-bsm2050-bedside-monitor-software-version-1-54-20. Based on FDA recall #Z-1169-2025 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1169-2025%22&limit=1).

The FDA announced on June 6, 2023, that Murata Vios, Inc. is recalling its Vios Monitoring System Bedside Monitor, Model BSM2050, with software versions 1.54.20 and 1.56.00 due to a premature failure in the L2050R pulse oximeter sensor. This defect causes the system to incorrectly display a 'Pulse Ox Sensor Off' warning. The FDA issued recall number Z-1169-2023.

How to tell if you have it

  • Brand: Murata Vios
  • Product: Vios Monitoring System Bedside Monitor, Model BSM2050
  • Software Versions: 1.54.20 and 1.56.00
  • UDI-DI: B595BSM20500
  • Quantity: 44 systems
  • Distribution: US Nationwide distribution in the state of NY

The hazard

A premature failure condition was identified with the L2050R pulse oximeter SpO2 sensor that connects to the L2050G Vios chest sensor device. When a failure occurs, a 'Pulse Ox Sensor Off' warning condition will be incorrectly displayed by the system.

Context

GovInfoNow's records show Murata Vios, Inc. has 3 other recalls on record since June 21, 2021. Other recent recalls of similar products include Philips IntelliVue MX40: Patient Monitor May Not Alarm - Recall (Oct. 31, 2025), Philips IntelliVue MX60: Monitor May Not Alarm - Recall (Oct. 31, 2025) and Philips IntelliVue MP60 Monitors May Not Alarm - Recall (Oct. 31, 2025).

How to tell if you have this one

Check for the model BSM2050 with software versions 1.54.20 or 1.56.00. The unique identifier (UDI-DI) is B595BSM20500.

  • Brand
    Murata Vios
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Check Software Version

    Look for the software version number on the device label or in the device settings.

Where it went and when

Where it was distributed

Distributed nationwide.

Alaska: yesAKMaine: yesMEVermont: yesVTNew Hampshire: yesNHWashington: yesWAIdaho: yesIDMontana: yesMTNorth Dakota: yesNDMinnesota: yesMNIllinois: yesILWisconsin: yesWIMichigan: yesMINew York: yesNYRhode Island: yesRIMassachusetts: yesMAOregon: yesORNevada: yesNVWyoming: yesWYSouth Dakota: yesSDIowa: yesIAIndiana: yesINOhio: yesOHPennsylvania: yesPANew Jersey: yesNJConnecticut: yesCTCalifornia: yesCAUtah: yesUTColorado: yesCONebraska: yesNEMissouri: yesMOKentucky: yesKYWest Virginia: yesWVVirginia: yesVAMaryland: yesMDDelaware: yesDEArizona: yesAZNew Mexico: yesNMKansas: yesKSArkansas: yesARTennessee: yesTNNorth Carolina: yesNCSouth Carolina: yesSCDistrict of Columbia: yesDCOklahoma: yesOKLouisiana: yesLAMississippi: yesMSAlabama: yesALGeorgia: yesGAHawaii: yesHITexas: yesTXFlorida: yesFL

FDA distribution: “US Nationwide distribution in the state of NY.” States read from FDA’s own wording; “not listed” means the notice does not name it.

Distributed nationwide.

Recall timeline

FDA published this recall 20 months after the firm started it.

1. Firm started the recall: Jun 6, 202312. FDA classified it Class II: Feb 12, 202523. Published in FDA’s Enforcement Report: Feb 19, 2025320232025
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report

Dates from the agency record only; steps without a date are left out.

✎ GovInfoNow

Questions people ask

Is there a risk of serious injury?

The recall record states the hazard is a false warning, which is unlikely to cause serious injury but could delay medical care.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-1169-2025 (June 6, 2023) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-1169-2025 to see it in context.
  • Plain-language summary written by qwen3:4b on August 19, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA