PTW-FREIBURG BeamAdjust Software 2.3.3: Incorrect Calibration Results Risk

PTW-FREIBURG recalls BeamAdjust software version 2.3.3 (Ref: S080032) for OCTAVIUS Detectors 1000 and 1600 models. Using this software with Tiff or DICOM files as reference may cause inaccurate measurements.

🇺🇸 FDA recall #Z-2577-2023Medical devicePTW-FREIBURG ›
Take action

Contact the company

Recalling firmPTW-FREIBURGLorracher Str. 7 · Freiburg Im Breisgau, Germany
✎ GovInfoNow plain-language explainer

PTW-FREIBURG is recalling BeamAdjust software version 2.3.3 for specific OCTAVIUS Detectors 1000 and 1600 models. This software version can produce incorrect calibration files when used with Tiff or DICOM data sets, leading to inaccurate measurement results. Read more at govinfonow.com/recall/57881-ptw-freiburg-beamadjust-software-2-3-3-incorrect-calibration-results.

Source Summary by GovInfoNow · govinfonow.com/recall/57881-ptw-freiburg-beamadjust-software-2-3-3-incorrect-calibration-results · Updated August 19, 2026 · FDA enforcement record: FDA #Z-2577-2023

Cite this page

“PTW-FREIBURG BeamAdjust Software 2.3.3: Incorrect Calibration Results Risk.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/57881-ptw-freiburg-beamadjust-software-2-3-3-incorrect-calibration-results. Based on FDA recall #Z-2577-2023 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2577-2023%22&limit=1).

The Food and Drug Administration announced on June 7, 2023, that PTW-FREIBURG is recalling BeamAdjust software Version 2.3.3 because it may generate incorrect calibration files for OCTAVIUS Detectors. This error causes measurement results to be inaccurate when using specific reference matrices.

How to tell if you have it

  • BeamAdjust software Version 2.3.3 Ref: S080032
  • OCTAVIUS Detector 1600 SRS (T10056) UDI-DI: EPTWT100560
  • OCTAVIUS Detector 1600 MR (T10057) UDI-DI: EPTWT100570
  • OCTAVIUS Detector 1600 XDR (T10058) UDI-DI: EPTWT100580
  • OCTAVIUS Detector 1000 SRS (T10036) UDI-DI: EPTWT100360

The hazard

According to the FDA, the software generates an incorrect calibration file if a relative calibration is performed using a Tiff file or DICOM data set as a reference matrix. Consequently, measurement results will be incorrect when using such a calibration file.

Context

GovInfoNow's records show PTW-FREIBURG has two other recalls on record since 2024-02-08. On 2026-08-04, the company recalled VERIQA RT MonteCarlo 3D v3 Software due to a hazard where incorrect dose delivery may occur if a bolus is omitted in treatment plans. Another recall on 2024-02-08 involved PTW-FREIBURG VERIQA Software Version 2.0/2.1, which risked false positive results by calculating incorrect Gamma Passing Rates for individual ROIs.

How to tell if you have this one

Check if your OCTAVIUS Detector 1000 SRS, 1600 SRS, 1600 MR, or 1600 XDR model has BeamAdjust software version 2.3.3. The software version can be identified by the UDI-DI codes listed in the recall notice.

  • Brand
    PTW-FREIBURG
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Update Software

    Install the corrected BeamAdjust software version from the manufacturer to ensure accurate calibration.

Where it went and when

Where it was distributed

Distributed nationwide.

Alaska: yesAKMaine: yesMEVermont: yesVTNew Hampshire: yesNHWashington: yesWAIdaho: yesIDMontana: yesMTNorth Dakota: yesNDMinnesota: yesMNIllinois: yesILWisconsin: yesWIMichigan: yesMINew York: yesNYRhode Island: yesRIMassachusetts: yesMAOregon: yesORNevada: yesNVWyoming: yesWYSouth Dakota: yesSDIowa: yesIAIndiana: yesINOhio: yesOHPennsylvania: yesPANew Jersey: yesNJConnecticut: yesCTCalifornia: yesCAUtah: yesUTColorado: yesCONebraska: yesNEMissouri: yesMOKentucky: yesKYWest Virginia: yesWVVirginia: yesVAMaryland: yesMDDelaware: yesDEArizona: yesAZNew Mexico: yesNMKansas: yesKSArkansas: yesARTennessee: yesTNNorth Carolina: yesNCSouth Carolina: yesSCDistrict of Columbia: yesDCOklahoma: yesOKLouisiana: yesLAMississippi: yesMSAlabama: yesALGeorgia: yesGAHawaii: yesHITexas: yesTXFlorida: yesFL

FDA distribution: “US Nationwide distribution.” States read from FDA’s own wording; “not listed” means the notice does not name it.

Distributed nationwide.

Recall timeline

FDA published this recall 3 months after the firm started it.

1. Firm started the recall: Jun 7, 202312. FDA classified it Class II: Sep 11, 202323. Published in FDA’s Enforcement Report: Sep 20, 202332023
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-2577-2023 (June 7, 2023) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-2577-2023 to see it in context.
  • Plain-language summary written by qwen3:4b on August 19, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA