Philips recalls Ingenia Elition X MRI systems with specific product numbers due to risk of bypassing smoke detector interlock after power cycling.
The system operator is able to bypass the SmokeDetector Interlock system locking mechanism after a smoke detection by power cycling the system.
Philips is recalling certain Ingenia Elition X MRI systems because users can bypass the smoke detector safety feature by restarting the machine. This could allow smoke to enter the system and cause fire or injury.
The system's smoke detector interlock can be disabled by power cycling the machine. This means smoke alarms may not activate during a fire, potentially allowing fire to spread inside the MRI room.
Medical facilities using the recalled MRI systems are most at risk. Owners of these specific product numbers should check their equipment.
Check for product numbers 781358, 782107, or 782118 on the MRI system. Serial numbers are listed in the detailed recall information provided by Philips.
Verify if your MRI system has the recalled product numbers or serial numbers listed in the official recall notice.
The system operator can bypass the smoke detector interlock by restarting the machine, which could allow fire to spread in the MRI room.
Check for product numbers 781358, 782107, or 782118 on the system. Serial numbers are provided in the detailed recall information.
No, but you should check if your system is affected and follow the recall instructions to ensure safety.
We’ve drafted a message you can send Philips to request your refund or repair — edit it as you like.
Subject: Recall Z-2395-2023 — Philips Philips Hello, I own a Philips that is covered by recall Z-2395-2023 from Philips. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.