Philips recalls specific MRI systems due to gradient coil overheating that may cause smoke or fire. Affected models include 781358, 782107, 782119, and 782136 with serial numbers listed in the notice.
Gradient Coil may act as a heat source with a potential to produce smoke and/or fire.
Philips is recalling certain Ingenia Elition X MRI systems because the gradient coil can overheat and potentially cause smoke or fire. This risk means affected units should be stopped from use immediately to prevent potential fire hazards.
The gradient coil in these MRI systems may overheat when in operation, producing smoke or fire. This occurs due to a design flaw that allows the coil to act as a heat source under normal operating conditions.
Healthcare facilities using the recalled MRI systems are most at risk. Patients and staff at these facilities could face fire hazards if the device is used without proper safety measures.
Check for model numbers 781358, 782107, 782119, or 782136 on the device. Serial numbers listed in the recall notice include specific ranges like 45002-45429 for model 781358.
Discontinue use of the affected MRI system to prevent potential fire hazards.
Model numbers 781358, 782107, 782119, and 782136 with specific serial numbers listed in the recall notice.
Look for the model numbers and serial numbers listed in the recall notice, which include ranges like 45002-45429 for model 781358.
Stop using the device immediately and contact Philips for further instructions.
We’ve drafted a message you can send Philips to request your refund or repair — edit it as you like.
Subject: Recall Z-0460-2023 — Philips Philips Hello, I own a Philips that is covered by recall Z-0460-2023 from Philips. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.