The FDA announced on July 13, 2023, that Siemens Medical Solutions USA, Inc. is recalling ACUSON Redwood 2.0 ultrasound systems due to a software error causing underestimated measurements. This defect may lead to misdiagnosis or negative patient management decisions. The agency classified the recall as Class II.
How to tell if you have it
- Brand: Siemens
- Product: ACUSON Redwood 2.0 ultrasound systems
- REF: 11503314
- UDI-DI: 04056869251264
- Software version prefix: VA20
The hazard
If a user-generated preset for an 18L6 transducer created on a 1.0 ultrasound system is used with a 2.0 system, the device displays underestimated measurement results in Dual format visualization. This error may lead to misdiagnosis of a patient's condition or influence patient management decisions negatively.
Context
GovInfoNow's records show Siemens Medical Solutions USA, Inc. has 182 other recalls on record since June 6, 2017. Other recent recalls include the Siemens ARTIS icono Ceiling Monitor recall for loose screws (June 8, 2026) and the Siemens ARTIS Icono Biplane recall for movement loss risk (May 20, 2026). A March 12, 2026 recall of Siemens ARTIS Pheno VE30A/VE40A units cited risks of inaccurate 3D scans due to radiation exposure.