Hamilton Medical recalls T1 ventilators with serial numbers 15929-19490 due to capacitors that may leak electrolyte, causing short circuits and ventilation interruptions. Affected units must be checked for proper function.
Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.
Hamilton Medical is recalling certain T1 ventilators because capacitors on the control board can leak electrolyte, potentially causing short circuits or ventilation interruptions. This could lead to life-threatening situations if the ventilator stops working during critical care.
Degrading capacitors in the ventilator's control board may leak electrolyte, causing short circuits that interrupt ventilation. This could lead to the ventilator stopping function during critical care, risking patient safety.
Healthcare professionals using the Hamilton T1 ventilator for intensive care or transport are most at risk. Patients on ventilators with affected units could experience interrupted breathing.
Check the serial number on the ventilator's label. Affected units have serial numbers listed in the recall: 15929, 17600, 17661, and other numbers up to 19490.
Ensure the ventilator is operating correctly; if it stops working unexpectedly, stop using it immediately.
Reach out to Hamilton Medical for guidance on whether your unit is affected and next steps.
Degrading capacitors on the control board may leak electrolyte, causing short circuits and ventilation interruptions.
Check the serial number on the ventilator's label. Affected units have serial numbers listed in the recall: 15929, 17600, 17661, and other numbers up to 19490.
Stop using it immediately and contact Hamilton Medical for assistance.
We’ve drafted a message you can send Hamilton Medical to request your refund or repair — edit it as you like.
Subject: Recall Z-2526-2023 — Hamilton Medical Hamilton Medical Hello, I own a Hamilton Medical that is covered by recall Z-2526-2023 from Hamilton Medical. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.