Hamilton Medical recalls control board spare parts for ventilators (model C1/T1/MR1) due to potential electrolyte leaks causing short circuits and ventilation interruptions. Affected units include specific serial numbers and reference codes.
Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.
Hamilton Medical is recalling spare control board parts for ventilators (C1/T1/MR1 models) because faulty capacitors may leak electrolyte, causing short circuits or ventilation interruptions. This could lead to life-threatening situations if the ventilator stops working unexpectedly.
Degrading capacitors in the spare control board parts may leak electrolyte, causing short circuits that interrupt ventilation. This could lead to sudden loss of ventilator function, risking patient safety.
Healthcare facilities using Hamilton C1/T1/MR1 ventilators with the affected spare parts (serial numbers 4001901, 4002139, etc.) are most at risk. Patients relying on these ventilators could face life-threatening interruptions.
Check if your ventilator model is C1/T1/MR1 and the spare control board part has the serial numbers listed: 4001901, 4002139, 4003445, 4003567, 4003576, 4005595. The reference code MSP161502/12 also identifies these parts.
Immediately stop using the affected spare control board parts (serial numbers 4001901, 4002139, 4003445, 4003567, 4003576, 4005595) as they may cause ventilation interruptions.
Hamilton Medical is recalling spare control board parts for ventilators (C1/T1/MR1 models) with the serial numbers 4001901, 4002139, 4003445, 4003567, 4003576, and 4005595.
Check the serial number on the spare control board part; affected units have serial numbers 4001901, 4002139, 4003445, 4003567, 4003576, or 4005595.
Immediately stop using the affected spare parts and contact Hamilton Medical for replacement or repair instructions.
We’ve drafted a message you can send Hamilton Medical to request your refund or repair — edit it as you like.
Subject: Recall Z-2529-2023 — Hamilton Medical Hamilton Medical Hello, I own a Hamilton Medical that is covered by recall Z-2529-2023 from Hamilton Medical. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.