Hamilton Medical recalls Hamilton-T1 ventilators with serial numbers listed in the record due to capacitors that may leak electrolyte, causing short circuits and ventilation interruptions.
Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.
Hamilton Medical is recalling certain Hamilton-T1 intensive care and transport ventilators because capacitors on the control board may leak electrolyte, leading to short circuits or loss of function that could interrupt ventilation. This is a serious issue for medical devices that must operate reliably.
Degrading capacitors on the control board may leak electrolyte, causing short circuits in the device. This can lead to the capacitor losing function and interrupting ventilation, which is critical for patients who depend on the ventilator for breathing.
Healthcare facilities using the affected Hamilton-T1 ventilators are most at risk. Patients relying on these ventilators for critical care could face interrupted breathing support if the device fails.
Check if your ventilator has a serial number listed in the provided range (e.g., 500816, 501879, etc.). The product is identified by the Hamilton-T1 model and reference number 161009.
Verify your ventilator's serial number against the list of affected numbers provided in the recall notice.
Reach out to Hamilton Medical for guidance on next steps if your ventilator is affected.
The Hamilton-T1 intensive care and transport ventilator with reference number 161009.
Check the serial number against the list of affected numbers provided in the recall notice.
Contact Hamilton Medical for guidance on next steps.
We’ve drafted a message you can send Hamilton Medical to request your refund or repair — edit it as you like.
Subject: Recall Z-2527-2023 — Hamilton Medical Hamilton Medical Hello, I own a Hamilton Medical that is covered by recall Z-2527-2023 from Hamilton Medical. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.