Puritan Bennett Cuff Pressure Manager REF 180-03: Power Port May Detach, Causing Treatment

Puritan Bennett recalls 180-03 Cuff Pressure Manager units with serial numbers E223500003-E225000099. Inadequate internal fixation of the power port may cause detachment after use, preventing charging and risking treatment delays.

🇺🇸 FDA recall #Z-2651-2023Medical devicePuritan Bennett ›
Take action

Contact the company

Recalling firmCovidien6135 Gunbarrel Ave · Boulder, CO 80301-3214
✎ GovInfoNow plain-language explainer

Puritan Bennett is recalling Cuff Pressure Manager devices with serial numbers starting with E223500003 to E225000099. The power port may become loose after several uses, preventing the device from charging and potentially delaying medical treatment. Read more at govinfonow.com/recall/57457-puritan-bennett-cuff-pressure-manager-ref-180-03-power-port.

Source Summary by GovInfoNow · govinfonow.com/recall/57457-puritan-bennett-cuff-pressure-manager-ref-180-03-power-port · Updated August 19, 2026 · FDA enforcement record: FDA #Z-2651-2023

Cite this page

“Puritan Bennett Cuff Pressure Manager REF 180-03: Power Port May Detach, Causing Treatment.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/57457-puritan-bennett-cuff-pressure-manager-ref-180-03-power-port. Based on FDA recall #Z-2651-2023 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2651-2023%22&limit=1).

The FDA announced on Aug. 22, 2023, that Covidien is voluntarily recalling Puritan Bennett Cuff Pressure Manager devices due to a hazard that could lead to treatment delay. Owners should contact the company for a remedy.

How to tell if you have it

  • Brand: Puritan Bennett
  • Product: Cuff Pressure Manager, REF: 180-03
  • UDI-DI: 10884521782495
  • Serial Numbers: E223500003 through E223500032, E223500054, E223500055, E223500079 through E223500083, E223600009 through E223600024, E223600042, E223600070 through E223600100, E224900001 through E224900015, E224900017 through E224900020, E224900021 through E224900026, E224900027 through E224900030, E224900031 through E224900037, E224900039 through E224900045, E224900047 through E224900050, E224900052, E224900053, E224900055 through E224900062, E224900064 through E224900070, E224900072 through E224900078, E224900081 through E224900085, E224900087, E224900091 through E224900095, E224900098, E224900099, E224900100, E225000001 through E225000007, E225000011 through E225000026, E225000027, E225000029, E225000031 through E225000040, E225000041 through E225000050, E225000051 through E225000060, E225000061 through E225000070, E225000071, E225000072, E225000073, E225000075 through E225000080, E225000081 through E225000090, E225000091 through E225000100, E230200001 through E230200010, E230200011 through E230200020, E230200021 through E230200030, E230200031 through E230200040, E230200041 through E230200050, E230200051 through E230200060, E230200061, E230200062, E230200064 through E230200070, E230200071 through E230200080, E230200081 through E230200090, E230200091, E230200092, E230200095, E230200096, E230500006 through E230500011, E230500031 through E230500040, E230500041 through E230500050, E230500051 through E230500060, E230500061 through E230500066, E230500092 through E230500094, E230500103 through E230500112, E230900079, E230900080

The hazard

Inadequate internal fixation of the power inlet port to the cuff pressure manager device causes the external power inlet to become dislodged within the device housing after several uses. This prevents the device from being charged and could lead to treatment delay.

Context

GovInfoNow's records show Covidien has 14 other recalls on record since March 5, 2021. Recent recalls include a June 23, 2025, Bedside SpO2 Monitor recall for alarms that may not be heard and a May 21, 2025, Newport HT70 Ventilator recall for risk of shutdown or alarm failure. Other recent recalls of similar products include the Hamilton Medical IntelliCuff Recall (May 22, 2026) and the COAPTIUM CONNECT Devices recall (June 15, 2026).

How to tell if you have this one

Check the serial number on the device. The affected units have serial numbers starting with E223500003 through E225000099. The product is sold under the brand name Puritan Bennett.

  • Brand
    Puritan Bennett
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Check Device Serial Number

    Look for the serial number on the device, which starts with E223500003 through E225000099.

  2. Stop Using if Detached

    If the power port becomes loose or detached, stop using the device immediately to prevent treatment delays.

Where it went and when

Where it was distributed

Distributed in 16 states, and outside the U.S.

Alaska: not listedAKMaine: not listedMEVermont: not listedVTNew Hampshire: not listedNHWashington: not listedWAIdaho: not listedIDMontana: not listedMTNorth Dakota: not listedNDMinnesota: not listedMNIllinois: yesILWisconsin: not listedWIMichigan: yesMINew York: yesNYRhode Island: not listedRIMassachusetts: yesMAOregon: not listedORNevada: not listedNVWyoming: not listedWYSouth Dakota: not listedSDIowa: not listedIAIndiana: not listedINOhio: not listedOHPennsylvania: not listedPANew Jersey: yesNJConnecticut: not listedCTCalifornia: yesCAUtah: yesUTColorado: yesCONebraska: not listedNEMissouri: not listedMOKentucky: yesKYWest Virginia: not listedWVVirginia: not listedVAMaryland: not listedMDDelaware: not listedDEArizona: yesAZNew Mexico: not listedNMKansas: yesKSArkansas: not listedARTennessee: yesTNNorth Carolina: yesNCSouth Carolina: not listedSCDistrict of Columbia: not listedDCOklahoma: not listedOKLouisiana: not listedLAMississippi: not listedMSAlabama: yesALGeorgia: not listedGAHawaii: not listedHITexas: yesTXFlorida: yesFL

FDA distribution: “US: AL, NJ, MA, MI, NY, NC, CO, TX, FL, KY, UT, CA, AZ, KS, IL, TN. OUS: El Salvador, Sweden, Canary Islands” States read from FDA’s own wording; “not listed” means the notice does not name it.

States: Alabama, Arizona, California, Colorado, Florida, Illinois, Kansas, Kentucky, Massachusetts, Michigan, New Jersey, New York, North Carolina, Tennessee, Texas, Utah.

Recall timeline

FDA published this recall 50 days after the firm started it.

1. Firm started the recall: Aug 22, 202312. FDA classified it Class II: Sep 29, 202323. Published in FDA’s Enforcement Report: Oct 11, 202332023
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-2651-2023 (August 22, 2023) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-2651-2023 to see it in context.
  • Plain-language summary written by qwen3:4b on August 19, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA