CareFusion 303, Inc. recalled BD Pyxis CII Safe ES automated dispensing cabinets and related desktop PCs due to a software error that could allow diversion of controlled substances, the FDA said on Sept. 6, 2023.
How to tell if you have it
- BD Pyxis CII Safe ES, REF: 1116-00
- BD Pyxis : CII Safe ES Desktop PC, Wired SCNR, REF: 107-254-01; CII Safe ES Desktop PC, W-LESS SCNR, BIO, REF: 107-255-01
- Enterprise Server, REF: 1115-00, 134157-01, 137670-02, 134148-01, 135292-01, 136449-01, 133760-01; REF: 1116-00
- UDI-DI: 10885403512605
- Software Versions: 1.7.3 and 1.7.4
The hazard
The FDA reported that when global edit is used to update multiple formulary properties simultaneously, specific settings are overwritten to "NO" without user warning. This error could result in the diversion of controlled substances from the automated dispensing cabinet.
Context
GovInfoNow's records show CareFusion 303, Inc. has 64 other recalls on record since 2017-06-12. Other recent recalls include BD Pyxis MedStation ES REF 323 Medication Cabinet Recall (2026-04-30), CareFusion labeling recall (2026-02-11), and BD Pyxis ES Enterprise Server Software Update Issue Recall (2025-12-03). Other recent recalls of similar products include Fresenius Kabi Ivenix Infusion System Software Recall (2025-11-14), B Braun IV Sets Risk Backflow of Medication - Recall (2025-10-29), and Elekta MOSAIQ Oncology System Recall (2025-12-10).