The FDA announced on September 21, 2023, that Acclarent, Inc. is recalling the TruDi Navigation System, Model FG-2000-00, Catalog ENS022B, when used with specific TruDi Curette models due to a location discrepancy hazard. This recall affects 141 units distributed across multiple U.S. states.
How to tell if you have it
- Brand: Acclarent
- Product: TruDi Navigation System, Model FG-2000-00, Catalog ENS022B
- Compatible Curette Models: TDC0005Z and TDC0005
- Software Version: V2.3.1 Update (2.3.1.144 and 2.3.1.166)
- UDI-DI: 10846835018639
- Quantity: 141 units
The hazard
According to the FDA, a discrepancy exists between the actual curette tip location and the location displayed on the navigation system during ENT and skull base surgery. This error may cause delayed or prolonged surgery, cerebrospinal fluid leak, visual impairment, or skull base structural damage.
Context
GovInfoNow's records show Acclarent, Inc. has two other recalls on record since March 6, 2019. On November 12, 2021, the company recalled TruDi NAV Suction Instruments due to calibration errors that risked tissue damage. Other recent recalls of similar products include the Encore Medical EMPOWR Acetabular System Recall (January 5, 2026) and the GE Medical Systems Imactis CT-Navigation System Recall (February 3, 2026).