Skip to content
Lower risk Medical device recall · FDA · Published September 28, 2023

Abbott NT2000iX RF Generator Recall: Potential Capacitance Failure Risk

Abbott recalls NT2000iX RF generators with specific serial numbers due to potential capacitance failure from improper servicing. Affected units may not provide accurate results.

🇺🇸 FDA recall #Z-0209-2024Medical deviceAbbott ›
Take action

Contact the company

Recalling firmAbbott5050 Nathan Ln N · Plymouth, MN 55442-3209
✎ GovInfoNow plain-language explainer

What happened, in plain English

Abbott is recalling certain NT2000iX radiofrequency generators that could have capacitance failure due to improper servicing. This may lead to inaccurate results during medical procedures. Read more at govinfonow.com/recall/57217-abbott-nt2000ix-rf-generator-recall-potential-capacitance-failure-risk.

Source Summary by GovInfoNow · govinfonow.com/recall/57217-abbott-nt2000ix-rf-generator-recall-potential-capacitance-failure-risk · Updated August 19, 2026 · FDA enforcement record: FDA #Z-0209-2024

Cite this page

“Abbott NT2000iX RF Generator Recall: Potential Capacitance Failure Risk.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/57217-abbott-nt2000ix-rf-generator-recall-potential-capacitance-failure-risk. Based on FDA recall #Z-0209-2024 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0209-2024%22&limit=1).

Why it’s dangerous

The hazard occurs when the generator is serviced using a tool that is out of tolerance, which can cause inaccurate capacitance measurements. This may lead to potential capacitance failure in the device, but does not cause immediate harm during normal use.

Who’s affected

Medical professionals using Abbott NT2000iX RF generators for lesioning procedures are most at risk if the device has been serviced with an out-of-tolerance tool. Patients using these devices are not directly affected.

How to tell if you have this one

Check for serial numbers listed in the recall: 16930509, 16983330, 17583407, and other specific numbers provided in the notice. The device is a radiofrequency generator used for medical lesioning procedures.

  • Brand
    Abbott
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Check Serial Numbers

    Verify if your device has a serial number from the list provided in the recall notice.

How this recall unfolded

Recall timeline

FDA published this recall 41 days after the firm started it.

1. Firm started the recall: Sep 28, 202312. FDA classified it Class II: Nov 2, 202323. Published in FDA’s Enforcement Report: Nov 8, 202332023
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report

Dates from the agency record only; steps without a date are left out.

✎ GovInfoNow

Questions people ask

Do I need to stop using the device immediately?

No, the recall does not require immediate discontinuation; affected users should check their serial number and follow the provided steps.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-0209-2024 (September 28, 2023) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-0209-2024 to see it in context.
  • Plain-language summary written by qwen3:4b on August 19, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA