CareFusion recalls MEDFLEX and MEDBANK automated dispensing cabinets that could dispense medication without pharmacist validation if 'Clear All' is selected on patient profiles.
If "Clear All" selected medications is selected on patient profile, the automated dispensing cabinet may allow dispensing of medication without first requiring the RxVerify Validation Code be entered/without validation from a pharmacist, which could lead to unintended and incorrect medication removal.
CareFusion is recalling several automated medication dispensing cabinets (MEDFLEX and MEDBANK models) because a software bug might allow dispensing medication without pharmacist validation. This could lead to incorrect medications being dispensed to patients.
If the 'Clear All' option is selected on a patient's profile in the cabinet's software, the system may dispense medication without first requiring the RxVerify Validation Code from a pharmacist. This could result in unintended and incorrect medications being dispensed to patients.
Healthcare facilities using CareFusion MEDFLEX or MEDBANK cabinets with affected software versions. Patients receiving medication from these cabinets may be at risk of receiving incorrect medications.
Check if your cabinet has a model name like MEDFLEX, MEDBANK, or MEDBANK MINI with the specific reference numbers listed in the recall notice. Affected software versions include 2.1.2.11, 2.2.1.8, 2.31.8, 2.3.2.1, and 2.4.2.26.
Verify your cabinet's software version by checking the device settings or documentation. Affected versions include 2.1.2.11, 2.2.1.8, 2.31.8, 2.3.2.1, and 2.4.2.26.
Reach out to CareFusion directly to confirm if your cabinet is affected and to get instructions for remediation.
Affected software versions include 2.1.2.11, 2.2.1.8, 2.31.8, 2.3.2.1, and 2.4.2.26.
Check your cabinet's model name and reference numbers against the list provided in the recall notice. Affected models include MEDFLEX, MEDBANK, and MEDBANK MINI with specific reference numbers.
No, but you should check your software version and contact CareFusion to determine if your cabinet is affected. The recall does not require immediate discontinuation of use.
We’ve drafted a message you can send CareFusion to request your refund or repair — edit it as you like.
Subject: Recall Z-1552-2025 — CareFusion CareFusion Hello, I own a CareFusion that is covered by recall Z-1552-2025 from CareFusion. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.