The FDA announced on 2023-10-17 that BioFire Diagnostics, LLC is recalling the BIOFIRE Blood Culture Identification 2 (BCID2) Panel due to potential false negative results for C. tropicalis infections. The agency stated that quality control testing may need amendment because affected control panels can lead to the pathogen being reported as not detected. Owners should contact BioFire Diagnostics for further instructions.
How to tell if you have it
- Brand: BioFire
- Product: BIOFIRE Blood Culture Identification 2 (BCID2) Panel, REF: RFIT-ASY-0147
- Used with: Maine Molecular Quality Controls, Inc (MMQCI) FilmArray BCID2 Control Panel M416
- Control Panel Reference: FLM1-PRT-0251; REF/UDI-DI: RFIT-ASY-0147/00815381020338
- Lot Code: FLM1-PRT-0251-16 and prior
- Units Affected: 32,910
The hazard
According to the FDA, routine quality control testing of the affected blood culture identification panel with the affected control panel may need to be amended. This is due to control panel and instrument characteristic changes that cause C. tropicalis to be amplified in some test runs but reported as "Not Detected" because the assay amplicon Tm is outside the acceptable limit for calling positive results.
Context
GovInfoNow's records show BioFire Diagnostics, LLC has 32 other recalls on record since 2018-03-14. Recent BioFire recalls include the Respiratory Panel 2.1 recall on 2026-06-24 for false negatives, the Joint Infection Panel Recall on 2026-04-22 for false positives, and the Spotfire Respiratory Panel Recall on 2026-04-01 for false negatives. Other recent recalls of similar products include the Biofire Defense FilmArray NGDS Warrior Panel Recall (2026-07-16), the Meridian Bioscience Revogene C. difficile Test Recall (2025-12-05), and the TECHNO-PATH Multichem ID-B Recall (2025-10-17).