The FDA announced on November 23, 2023, that FUJIFILM Healthcare Americas Corporation is recalling its mobile x-ray system due to incorrect exposure settings. The agency issued recall number Z-3158-2024 for the device.
How to tell if you have it
- Brand: FUJIFILM Healthcare
- Product: Mobile x-ray system designed to work with Fujifilm's GOS and CsI scintillator FDR D-EVO2 (DR-ID) 12XXSE family of digital x-ray detectors
- Quantity: 130 units
The hazard
According to the FDA, when the key switch is turned on, exposure conditions default to values intended for adult chest imaging. Additionally, if exposure conditions are set from the console while the key switch is off, those settings will not be reflected.
Context
GovInfoNow's records show FUJIFILM Healthcare Americas Corporation has 13 other recalls on record since October 14, 2021. Recent recalls include a July 22, 2026 recall for inaccurate ruler measurements in Synapse PACS Software v7.4.310 and a January 9, 2026 recall regarding overheating risks in the FDR Visionary Suite CH-200 X-ray Device. Other recent recalls of similar products include a June 5, 2026 column safety risk for the GE Optima XR200amx Mobile Radiography System and a November 18, 2025 recall for exceeding air kerma limits in the AGFA DR 800 X-Ray System.