Use caution Medical device recall · FDA · Published December 7, 2023

Philips Incisive CT X-Ray System Recall: PIM Cable May Cause ECG Signal Loss

Philips recalls Incisive CT models 728143/728144 with specific PIM cable part numbers due to risk of ECG signal loss during emergencies. Affected units may require replacement of the cable to prevent delays in diagnosis.

🇺🇸 FDA recall #Z-0719-2024Medical devicePhilips ›
Take action

Contact the company

Recalling firmPhilips North America222 Jacobs St · Cambridge, MA 02141-2296
✎ GovInfoNow plain-language explainer

Philips is recalling two specific models of their Incisive CT X-Ray systems because the PIM cable can become disconnected during frequent use, potentially causing ECG signal loss. This could delay critical diagnoses during emergencies, so affected users should check their cable part numbers and take action. Read more at govinfonow.com/recall/56599-philips-incisive-ct-x-ray-system-recall-pim-cable-may.

Source Summary by GovInfoNow · govinfonow.com/recall/56599-philips-incisive-ct-x-ray-system-recall-pim-cable-may · Updated August 19, 2026 · FDA enforcement record: FDA #Z-0719-2024

Cite this page

“Philips Incisive CT X-Ray System Recall: PIM Cable May Cause ECG Signal Loss.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/56599-philips-incisive-ct-x-ray-system-recall-pim-cable-may. Based on FDA recall #Z-0719-2024 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0719-2024%22&limit=1).

The FDA announced on Dec. 7, 2023, that Philips North America is recalling Incisive CT systems with specific PIM cables due to potential signal loss. The hazard involves intermittent ECG signal failure caused by cable disconnection during daily operation.

How to tell if you have it

  • Brand: Philips
  • Product: Incisive CT Computed Tomography X-Ray System with PIM cable option
  • Model Numbers: 728144 and 728143
  • PIM Cable Part Number: 459801179871
  • REF 728143 Serial Numbers: 33007, 33038, 33013, 33006, 33012, 33045, 33011, 500102, 33015, 33005, 33041, 33047, 500217, 500255, 33035, 500252, 500145, 500046, 500168, 33009, 33046, 33024, 500222, 500064, 33034, 500242, 500301, 33010, 33019, 33001, 33020, 33031, 33037, 500110, 500111, 500178, 33003, 500282, 33002, 33033
  • REF 728144 Serial Numbers: 34002, 34051, 34050, 34041, 34059, 34057, 34054, 34040, 34007, 34026, 34058, 34049, 34056, 34052
  • Distribution: Sold in AL, CA, CO, FL, GA, HI, IL, IN, KS, KY, LA, MD, MI, MO, MS, ND, NE, NJ, NY, OH, OK, PA, RI, SC, TX, UT, VT, WA and worldwide

The hazard

If the Patient Interface Monitor (PIM) Cable is plugged or unplugged frequently during daily operation, the connector and wire may disconnect from the welding point. This may result in intermittent loss of the ECG signal, which could lead to decisions to rescan patients or delay diagnosis during a clinical emergency.

Context

GovInfoNow's records show Philips North America has issued 30 other recalls since Dec. 24, 2020. Recent brand recalls include the Aug. 28, 2026, recall of Ingenia Elition S and X models for incorrect diffusion data, the July 31, 2026, recall of PIC iX Essentials Software for freezing issues, and the April 14, 2026, recall regarding MR 7700 MRE device stiffness errors. Other recent recalls of similar products include the Dec. 3, 2025, Philips Incisive CT recall for unstable sensors, the April 20, 2026, GE Medical Systems Allia IGS 3 Pulse X-Ray System recall for electrical shock risk, and the March 3, 2026, Philips Azurion 5 M12 X-ray System recall for startup failures.

How to tell if you have this one

Check the PIM cable part number on your ECG monitor connection to the Incisive CT gantry. Affected units have part number 459801179871. Model numbers 728143 and 728144 with specific serial numbers listed in the recall notice are affected.

  • Brand
    Philips
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Check PIM Cable Part Number

    Locate the connection cable between your ECG monitor and Incisive CT gantry and verify the part number matches 459801179871.

Where it went and when

Where it was distributed

Distributed in 28 states, and outside the U.S.

Alaska: not listedAKMaine: not listedMEVermont: yesVTNew Hampshire: not listedNHWashington: yesWAIdaho: not listedIDMontana: not listedMTNorth Dakota: yesNDMinnesota: not listedMNIllinois: yesILWisconsin: not listedWIMichigan: yesMINew York: yesNYRhode Island: yesRIMassachusetts: not listedMAOregon: not listedORNevada: not listedNVWyoming: not listedWYSouth Dakota: not listedSDIowa: not listedIAIndiana: yesINOhio: yesOHPennsylvania: yesPANew Jersey: yesNJConnecticut: not listedCTCalifornia: yesCAUtah: yesUTColorado: yesCONebraska: yesNEMissouri: yesMOKentucky: yesKYWest Virginia: not listedWVVirginia: not listedVAMaryland: yesMDDelaware: not listedDEArizona: not listedAZNew Mexico: not listedNMKansas: yesKSArkansas: not listedARTennessee: not listedTNNorth Carolina: not listedNCSouth Carolina: yesSCDistrict of Columbia: not listedDCOklahoma: yesOKLouisiana: yesLAMississippi: yesMSAlabama: yesALGeorgia: yesGAHawaii: yesHITexas: yesTXFlorida: yesFL

FDA distribution: “Domestic distribution to the following states: AL CA CO FL GA HI IL IN KS KY LA MD MI MO MS ND NE NJ NY OH OK PA RI SC TX UT VT WA International distribution worldwide.” States read from FDA’s own wording; “not listed” means the notice does not name it.

States: Alabama, California, Colorado, Florida, Georgia, Hawaii, Illinois, Indiana, Kansas, Kentucky, Louisiana, Maryland, Michigan, Mississippi, Missouri, Nebraska, New Jersey, New York, North Dakota, Ohio, Oklahoma, Pennsylvania, Rhode Island, South Carolina, Texas, Utah, Vermont, Washington.

Recall timeline

FDA published this recall 48 days after the firm started it.

1. Firm started the recall: Dec 7, 202312. FDA classified it Class II: Jan 12, 202423. Published in FDA’s Enforcement Report: Jan 24, 2024320232024
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-0719-2024 (December 7, 2023) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-0719-2024 to see it in context.
  • Plain-language summary written by qwen3:4b on August 19, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA