Philips recalls ProxiDiagnost upgrade A Multi-functional Radiography and Fluoroscopy systems due to potential incorrect patient data display and security vulnerabilities in wireless detectors.
Issue 1: Mixed Images: The Radio Fluoroscopy (RF) viewer may also display a previous patients radiography (RAD) images when starting the next patient scan while the previous patient image export is still processing. If the issue occurs, there is a potential for incorrect patient data to be displayed in the image. Issue 2: Detector Access point: Security vulnerability specific to the Wireless Portable Detector configuration items in Philips Support Connect (PSC). Due to this vulnerability it is possible, with physical access to the system and knowledge of specific settings, to modify and export data to removable media (example: USB)
Philips is recalling its ProxiDiagnost upgrade A Multi-functional Radiography and Fluoroscopy systems because they may display previous patient images during new scans, risking incorrect patient data. The systems also have a security vulnerability that could allow data export to removable media if physically accessed.
The system may show previous patient images during new scans, which could lead to incorrect patient data being displayed. Additionally, wireless detectors have a security vulnerability that might allow data to be exported to removable media if someone has physical access and knows specific settings.
Healthcare providers using Philips ProxiDiagnost systems with serial numbers starting with US: 4 5 6 7 8 9 11 12 13 14 17 18 61 692222229, 692222254, or 692222263. Medical staff handling these systems are most at risk due to potential data errors and security issues.
Check for serial numbers starting with US: 4 5 6 7 8 9 11 12 13 14 17 18 61 692222229, 692222254, or 692222263. The system is sold under the Philips ProxiDiagnost upgrade A Multi-functional Radiography and Fluoroscopy (R/F) name.
Verify your system's serial number starts with US: 4 5 6 7 8 9 11 12 13 14 17 18 61 692222229, 692222254, or 692222263.
The security vulnerability affects wireless detectors and allows data to be exported to removable media if someone has physical access and knows specific settings.
Check for serial numbers starting with US: 4 5 6 7 8 9 11 12 13 14 17 18 61 692222229, 692222254, or 692222263.
If your serial number does not match the affected numbers, no action is required.
We’ve drafted a message you can send Philips to request your refund or repair — edit it as you like.
Subject: Recall Z-1073-2024 — Philips Philips Hello, I own a Philips that is covered by recall Z-1073-2024 from Philips. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.