Skip to content
Use caution Medical device recall · FDA · Published January 3, 2024

FUJIFILM Synapse Cardiology PACS Software Recalled: Risk of Misdiagnosis

FUJIFILM Healthcare recalls specific versions of Synapse Cardiology PACS software due to potential incorrect left ventricle mass calculations, which could lead to misdiagnosis or incorrect treatment.

🇺🇸 FDA recall #Z-1171-2024Medical deviceFUJIFILM Healthcare ›
Take action

Contact the company

Recalling firmFUJIFILM Healthcare Americas Corporation81 Hartwell Ave Ste 300 · Lexington, MA 02421-3160
✎ GovInfoNow plain-language explainer

FUJIFILM Healthcare is recalling specific versions of their Synapse Cardiology PACS software. The software may calculate left ventricle mass incorrectly, which could lead to misdiagnosis or incorrect treatment plans and serious health consequences. Read more at govinfonow.com/recall/56415-fujifilm-synapse-cardiology-pacs-software-recalled-risk-of-misdiagnosis.

Source Summary by GovInfoNow · govinfonow.com/recall/56415-fujifilm-synapse-cardiology-pacs-software-recalled-risk-of-misdiagnosis · Updated August 19, 2026 · FDA enforcement record: FDA #Z-1171-2024

Cite this page

“FUJIFILM Synapse Cardiology PACS Software Recalled: Risk of Misdiagnosis.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/56415-fujifilm-synapse-cardiology-pacs-software-recalled-risk-of-misdiagnosis. Based on FDA recall #Z-1171-2024 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1171-2024%22&limit=1).

The FDA announced on Jan. 3, 2024, that FUJIFILM Healthcare Americas Corporation is recalling Synapse Cardiology PACS V7.3.0, V7.2, V7.1, and V7.0 due to a risk of misdiagnosis. The web-based application may calculate left ventricular mass incorrectly, potentially leading to serious health consequences for patients.

How to tell if you have it

  • Brand: FUJIFILM Healthcare
  • Product: Synapse Cardiology PACS V7.3.0, V7.2, V7.1, V7.0
  • Software versions: 7.3.000, 7.2.200, 7.2.100, 7.2.000, 7.1.000, 7.0.000
  • UDI-DI: 854904006008
  • Quantity: 13 units

The hazard

According to the FDA, the LV Mass (2D Bullet) equation may be calculated incorrectly, resulting in a variance in the mass of the left ventricle. If used as a main factor for diagnosis, there is a risk of misdiagnosis or incorrect treatment plan of a patient, resulting in long-term health consequences or serious deterioration of health.

Context

GovInfoNow's records show FUJIFILM Healthcare Americas Corporation has 13 other recalls on record since Oct. 14, 2021. Other recent recalls of similar products include GE Healthcare Centricity PACS Software Recall (July 31, 2026), Merge PACS Workstation Software Inaccurate Measurements - Recall (July 22, 2026), and Philips Cardiac Workstation 7000: Inaccurate Heart Rate Data Risk (July 1, 2026).

How to tell if you have this one

Check if you have Synapse Cardiology PACS software versions 7.3.0, 7.2, 7.1, or 7.0. The software is a web-based application used for processing medical images.

  • Brand
    FUJIFILM Healthcare
  • Category
    Medical device
FDA recall event · 2 products

All 2 products in this recall

FDA lists each product line separately under one recall event. Check your product description, lot or code numbers against each row. Every row shares the hazard and remedy on this page.

  • Classification
    Class II
  • Status
    Ongoing
  • Distribution
    US domestic distribution nationwide. International distribution to Israel.
Recall numberProduct, codes & lotsQuantity
Z-1171-2024Synapse Cardiology PACS V7.3.0, V7.2, V7.1, V7.0. A web-based application as the primary user interface for the processing of medical images.; UDI-DI 854904006008 Software versions 7.3.000, 7.2.200, 7.2.100, 7.2.000, 7.1.000, 7.0.00013
Z-1172-2024Synapse CV 6. with AR. A web-based application as the primary user interface for the processing of medical images.; UDI-DI 00854904006015 Software versions 6.2.1, 6.2.0, 6.1.1, 6.1.0, 6.0.1, and 6.0.08
Based on the official FDA remedy

What to do — step by step

  1. Update Software

    Contact FUJIFILM Healthcare to get the latest software version that fixes the issue.

Where it went and when

Where it was distributed

Distributed nationwide, and outside the U.S.

Alaska: yesAKMaine: yesMEVermont: yesVTNew Hampshire: yesNHWashington: yesWAIdaho: yesIDMontana: yesMTNorth Dakota: yesNDMinnesota: yesMNIllinois: yesILWisconsin: yesWIMichigan: yesMINew York: yesNYRhode Island: yesRIMassachusetts: yesMAOregon: yesORNevada: yesNVWyoming: yesWYSouth Dakota: yesSDIowa: yesIAIndiana: yesINOhio: yesOHPennsylvania: yesPANew Jersey: yesNJConnecticut: yesCTCalifornia: yesCAUtah: yesUTColorado: yesCONebraska: yesNEMissouri: yesMOKentucky: yesKYWest Virginia: yesWVVirginia: yesVAMaryland: yesMDDelaware: yesDEArizona: yesAZNew Mexico: yesNMKansas: yesKSArkansas: yesARTennessee: yesTNNorth Carolina: yesNCSouth Carolina: yesSCDistrict of Columbia: yesDCOklahoma: yesOKLouisiana: yesLAMississippi: yesMSAlabama: yesALGeorgia: yesGAHawaii: yesHITexas: yesTXFlorida: yesFL

FDA distribution: “US domestic distribution nationwide. International distribution to Israel.” States read from FDA’s own wording; “not listed” means the notice does not name it.

Distributed nationwide, and outside the U.S.

Recall timeline

FDA published this recall 56 days after the firm started it.

1. Firm started the recall: Jan 3, 202412. FDA classified it Class II: Feb 21, 202423. Published in FDA’s Enforcement Report: Feb 28, 202432024
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-1171-2024 (January 3, 2024) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-1171-2024 to see it in context.
  • Plain-language summary written by qwen3:4b on August 19, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA