Ortho-Clinical Diagnostics, Inc. voluntarily recalled VITROS XT 7600 Integrated Systems with catalog number 6844461 running software versions 3.8.0 or 3.8.1 due to a software defect that may cause erroneous patient results, the FDA said on Jan. 26, 2024.
How to tell if you have it
- Brand: Ortho-Clinical
- Product: VITROS XT 7600 Integrated System
- Catalog No.: 6844461
- Software Versions: 3.8.0, 3.8.1
- UDI-DI: 10758750031610
- US Serial Numbers: J76000050 through J76000873
- OUS Serial Numbers: J76000106 through J76001761
The hazard
A software defect causes Configurable Westgard Rules to not report as expected when a new control is defined. This may result in the acceptance of QC results that do not meet acceptance criteria, which may cause erroneous patient results to be reported.
Context
GovInfoNow's records show Ortho-Clinical Diagnostics, Inc. has 53 other recalls on record since 2017-07-10. Other recent recalls of similar products include Ortho-Clinical LAC Slides Recall (2025-07-29), Ortho-Clinical VITROS Ca Slides Recall (2025-07-09), and Ortho-Clinical VITROS OP Reagent Gen 15 May Cause Inaccurate Test Results (2025-06-12). Other recent recalls of similar products include Beckman Coulter DxC 500i Modules with Software 1.5 Risk (2026-07-15), Beckman Coulter DxC 500 AU Modules with Software 1.5 Risk (2026-07-15), and Roche cobas pro integrated solutions recall (2026-01-20).