The FDA announced on Jan. 26, 2024, that Ortho-Clinical Diagnostics is recalling VITROS 5600 Integrated Systems due to a software defect that may cause erroneous patient results. The agency classified the recall as Class II.
How to tell if you have it
- Brand: Ortho-Clinical
- Product: VITROS 5600 Integrated System
- Catalog Nos.: 6802413 (regular), 6802915 (refurbished)
- Software Version: 3.8.1
- Regular UDI-DI: 10758750002740
- Refurbished UD-DI: 10758750007110
- Serial Numbers (US): J56001885, J56003484, J56001886, J56003541, J56001910, J56003570, J56002308, J56003583, J56002461, J56003646, J56002474, J56003669, J56002608, J56003670, J56002698, J56004068, J56002704, J56004137, J56003342, J56004174, J56003356, J56004491, J56003377, J56004506, J56003433, J56004549, J56003477, J56004551, J56003478, J56004577
- Serial Numbers (OUS): J56000425, J56002518, J56000512, J56002615, J56000533, J56002625, J56000674, J56002637, J56000680, J56002926, J56000719, J56002929, J56000809, J56003061, J56000831, J56003260, J56000836, J56003285, J56001051, J56003401, J56001111, J56003464, J56001114, J56003508, J56001208, J56003523, J56001731, J56003538, J56001765, J56004346, J56001826, J56004355, J56001853, J56004415, J56002085, J56004531, J56002116, J56004695
The hazard
A software defect in VITROS Systems running Software Version 3.8.1 causes QC baseline statistics to not update as expected when changed by the user. This may cause erroneous patient results to be reported.
Context
GovInfoNow's records show Ortho-Clinical Diagnostics, Inc. has 53 other recalls on record since 2017-07-10. Other recent recalls of similar products include Beckman Coulter DxC 500i Modules with Software 1.5 Risk: Erroneous Test Results (2026-07-15), Roche cobas pro integrated solutions recall: Software defect causes incorrect lab results (2026-01-20), and Beckman Coulter DxC 500 AU Modules with Software 1.5 Risk Patient Test Errors (2026-07-15).