The FDA announced on Feb. 7, 2024, that Siemens Healthcare Diagnostics is recalling DCA Systems HbA1c Reagent Kits due to a positive bias in test results. This defect may lead to incorrect patient management of glucose control.
How to tell if you have it
- Brand: Siemens Healthcare
- Product: DCA Systems HbA1c Reagent Kit
- Siemens Material Number: 10311134
- UDI/DI: 00630414532806
- Lot codes: 9640093, 0645103, 0648103, 0649103, 0654103, 0637093, 0641093, 0642093, 0647103, 0653103, 0655103
- Distribution: US Nationwide and Taiwan
The hazard
The recalled kits display a positive bias without alerting the user. The maximum bias observed was 19% when compared to NGSP pooled patient target-value assigned samples. Erroneous test results could potentially lead to incorrect patient management of glucose control.
Context
GovInfoNow's records show Siemens Healthcare Diagnostics has 150 other recalls on record since June 7, 2017. Other recent recalls include the Siemens Atellica CH RCRP Test recall from Aug. 17, 2026, and the Beckman Coulter Emit 2000 Quinidine Assay Recall from July 13, 2026. Similar product recalls include the Abbott Diabetes Care FreeStyle Libre 3 Plus Sensor Recall from Nov. 24, 2025.