The FDA announced on Feb. 12, 2024, that Boston Scientific Corporation is voluntarily recalling its POLARSHEATH Steerable Sheath 12F due to a manufacturing error that may cause the shaft to delaminate. The agency issued recall number Z-1302-2024 for the Class I device.
How to tell if you have it
- Brand: Boston Scientific
- Product: POLARSHEATH Steerable Sheath 12F
- REF: M004CRBS3050
- UPC: 08714729992684
- Batch Numbers: 32895945, 32895947, 32903274, 32903275, 32903276, 32903277, 32903278, 32903340, 32999267, 33000331, 33000333, 33000334, 33000337, 33000338, 33000339, 33052646, 33052647, 33052648, 33052649, 33061576, 33171976
The hazard
A tooling error in manufacturing may have caused delamination of the inner lumen of the sheath shaft in a subset of POLARSHEATH devices. This defect could result in the embolization of a fragment during use, such as when flushing the sheath or introducing a dilator or ablation catheter.
Context
GovInfoNow's records show Boston Scientific Corporation has 124 other recalls on record since June 1, 2017. Recent brand recalls include the Aug. 24, 2026, recall of Imager II Angiographic Catheters for tip fracture risk and the Aug. 6, 2026, recall of LUX-Dx Cardiac Monitor Systems for delayed alerts. Other recent recalls of similar products include the June 26, 2026, Boston Scientific Transcarotid Kit Sheath Depth Markings Recall, the Feb. 13, 2026, Merit Medical BioFlo DuraMax Catheter Recall, and the Dec. 22, 2025, Medline Livewire Steerable Catheters Recall.