Skip to content
Lower risk Medical device recall · FDA · Published February 29, 2024

BD Phoenix Yeast ID Panel Recall: Esculin Causes Test Cancellation

Becton Dickinson recalls BD Phoenix Yeast ID Panels due to esculin in A17 well causing system to abort tests. Affected units have specific UDI-DI and expiration dates.

🇺🇸 FDA recall #Z-1399-2024Medical deviceBecton Dickinson ›
Take action

Contact the company

Recalling firmBecton Dickinson & Co.7 Loveton Cir · Sparks, MD 21152-9212
✎ GovInfoNow plain-language explainer

Becton Dickinson is recalling BD Phoenix Yeast ID Panels that may contain esculin in the A17 well. This can cause the automated system to cancel tests, leading to inaccurate results. Users should check their panel's UDI-DI and expiration date to see if it's affected. Read more at govinfonow.com/recall/56031-bd-phoenix-yeast-id-panel-recall-esculin-causes-test-cancellation.

Source Summary by GovInfoNow · govinfonow.com/recall/56031-bd-phoenix-yeast-id-panel-recall-esculin-causes-test-cancellation · Updated August 19, 2026 · FDA enforcement record: FDA #Z-1399-2024

Cite this page

“BD Phoenix Yeast ID Panel Recall: Esculin Causes Test Cancellation.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/56031-bd-phoenix-yeast-id-panel-recall-esculin-causes-test-cancellation. Based on FDA recall #Z-1399-2024 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1399-2024%22&limit=1).

The FDA announced on Feb. 29, 2024, that Becton Dickinson & Co. is recalling BD Phoenix Yeast ID Panels due to a manufacturing defect. The hazard involves esculin contamination in an empty well, which can cause test cancellations. Owners should contact the company for details.

How to tell if you have it

  • BD Phoenix Yeast ID Panel
  • UPC: 30382904483167
  • Lot Number: 3312180
  • Expiration Date: 30-OCT-2024

The hazard

Esculin, a substrate normally found in the A16 well, was discovered in the A17 well, which is supposed to be empty. This contamination triggers fluorescence that causes the BD Phoenix Automated Microbiology System to abort the panel with a "Fluorescent interference has been detected" message.

Context

GovInfoNow's records show Becton Dickinson & Co. has 43 other recalls on record since 2017-06-01. Recent brand recalls include the BD GasPak EZ Campy Pouch System Recall (2026-04-07) and the BD Kiestra Isolate Suspension Cuvette Array recall (2025-10-23). Other recent recalls of similar products include Remel Yeastone Broth Recall (2026-03-24), Changchun Wancheng Bio-Electron UTI Test Strips Recall (2025-11-22), and Biomerieux VITEK 2 AST-N461 Test Kit Recall (2026-07-24).

How to tell if you have this one

Check the UDI-DI number on the panel: (10) 3312180 (17) 241030 (30)0025 (01) 30382904483167. The expiration date is 30-OCT-2024. This specific panel model is affected.

  • Brand
    Becton Dickinson
  • Product
    BD Phoenix Yeast ID Panel is a self-inoculating (when used with a BD Phoenix Automated Microbiology System) polystyrene
  • Category
    Medical device
FDA recall event · 2 products

All 2 products in this recall

FDA lists each product line separately under one recall event. Check your product description, lot or code numbers against each row. Every row shares the hazard and remedy on this page.

  • Classification
    Class II
  • Status
    Ongoing
  • Distribution
    Worldwide distribution - US Nationwide and the countries of AR, BE, CO, IN, JP, PE, PY, SG, TW, UY.
Recall numberProduct, codes & lotsQuantity
Z-1398-2024BD Phoenix NMIC-306 Panel is a self-inoculating (when used with a BD Phoenix Automated Microbiology System) polystyrene panel (tray) containing control wells and substrate wells with dried biochemical substrates that use traditional qualitative microbiology methods such as fermentation, oxidation, degradation, and hydrolysis in combination with chromogenic and fluorogenic substrates along with carbon sources for bacterial identification and quantitation.; UDI-DI: (10) 3318700 (17)241107 (30)0025 (01) 30382904492923 Lot Number: 3318700 Exp. Date: 07-NOV-202411,800 eaches
Z-1399-2024BD Phoenix Yeast ID Panel is a self-inoculating (when used with a BD Phoenix Automated Microbiology System) polystyrene panel (tray) containing control wells and substrate wells with dried biochemical substrates that use traditional qualitative microbiology methods such as fermentation, oxidation, degradation, and hydrolysis in combination with chromogenic and fluorogenic substrates along with carbon sources for yeast identification.; UDI-DI: (10) 3312180 (17) 241030 (30)0025 (01) 30382904483167 Lot Number: 3312180 Exp. Date: 30-OCT-20246700 eaches
Based on the official FDA remedy

What to do — step by step

  1. Check Panel Details

    Verify the UDI-DI number and expiration date on the panel to confirm if it's affected.

Where it went and when

Where it was distributed

Distributed nationwide, and outside the U.S.

Alaska: yesAKMaine: yesMEVermont: yesVTNew Hampshire: yesNHWashington: yesWAIdaho: yesIDMontana: yesMTNorth Dakota: yesNDMinnesota: yesMNIllinois: yesILWisconsin: yesWIMichigan: yesMINew York: yesNYRhode Island: yesRIMassachusetts: yesMAOregon: yesORNevada: yesNVWyoming: yesWYSouth Dakota: yesSDIowa: yesIAIndiana: yesINOhio: yesOHPennsylvania: yesPANew Jersey: yesNJConnecticut: yesCTCalifornia: yesCAUtah: yesUTColorado: yesCONebraska: yesNEMissouri: yesMOKentucky: yesKYWest Virginia: yesWVVirginia: yesVAMaryland: yesMDDelaware: yesDEArizona: yesAZNew Mexico: yesNMKansas: yesKSArkansas: yesARTennessee: yesTNNorth Carolina: yesNCSouth Carolina: yesSCDistrict of Columbia: yesDCOklahoma: yesOKLouisiana: yesLAMississippi: yesMSAlabama: yesALGeorgia: yesGAHawaii: yesHITexas: yesTXFlorida: yesFL

FDA distribution: “Worldwide distribution - US Nationwide and the countries of AR, BE, CO, IN, JP, PE, PY, SG, TW, UY.” States read from FDA’s own wording; “not listed” means the notice does not name it.

Distributed nationwide, and outside the U.S.

Recall timeline

FDA published this recall 41 days after the firm started it.

1. Firm started the recall: Feb 29, 202412. FDA classified it Class II: Mar 29, 202423. Published in FDA’s Enforcement Report: Apr 10, 202432024
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-1399-2024 (February 29, 2024) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-1399-2024 to see it in context.
  • Plain-language summary written by qwen3:4b on August 19, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA