The FDA announced on March 5, 2024, that Beckman Coulter, Inc. is recalling DxI 9000 Access Immunoassay Analyzers due to manufacturing defects that can cause leaking. This leak may lead to delays in reporting patient test results and exposure to physical and chemical hazards.
How to tell if you have it
- Brand: Beckman Coulter
- Product: DxI 9000 Access Immunoassay Analyzer, REF C11137
- Serial Numbers: 300254, 300256, 300257, 300258, 300259, 300263, 300264, 300266, 300268, 300269, 300270, 300271, 300272, 300273, 300275, 300281, and 300282
- Distribution: US Nationwide in New Hampshire; Belgium, Egypt, Italy, Poland, Spain
The hazard
Beckman Coulter manufacturing discovered cracks around a threaded insert on the bottom of the Internal Wash Buffer Tank used within the Wash Buffer Supply Drawer. This defect can result in leaking, leading to delays in reporting patient test results, exposure to physical and chemical hazards, and damage to equipment or the environment.
Context
GovInfoNow's records show Beckman Coulter, Inc. has 132 other recalls on record since June 20, 2017. Recent similar product recalls include the AU5800 Chemistry Analyzer Recall (Jan. 30, 2026) and the DxC 700 AU Recall (Jan. 30, 2026), both involving delays in results. Other recent recalls of similar products include the MicroScan LabPro V5.03 Panel Update 11 (June 26, 2026) and the Access Total T4 Calibrator Recall (March 16, 2026).