Skip to content
Unrated Medical device recall · FDA · Published March 6, 2024

Microbiologics KWIK-STIK Quality Control Kit Recall: Contaminated Fungal Control

Microbiologics recalls KWIK-STIK Rhizopus stolonifer control kits due to contamination with Aspergillus brasiliensis. Users may see A. brasiliensis growth. No remedy specified in recall notice.

🇺🇸 FDA recall #Z-1483-2024Medical deviceMicrobiologics ›
Take action

Contact the company

Recalling firmMicrobiologics Inc200 Cooper Ave N · Saint Cloud, MN 56303-4440
✎ GovInfoNow plain-language explainer

Microbiologics is recalling specific KWIK-STIK quality control kits because the fungal control material was contaminated with another fungus. This could cause users to mistakenly identify a different fungus when testing culture media. Read more at govinfonow.com/recall/55970-microbiologics-kwik-stik-quality-control-kit-recall-contaminated-fungal-control.

Source Summary by GovInfoNow · govinfonow.com/recall/55970-microbiologics-kwik-stik-quality-control-kit-recall-contaminated-fungal-control · Updated August 19, 2026 · FDA enforcement record: FDA #Z-1483-2024

Cite this page

“Microbiologics KWIK-STIK Quality Control Kit Recall: Contaminated Fungal Control.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/55970-microbiologics-kwik-stik-quality-control-kit-recall-contaminated-fungal-control. Based on FDA recall #Z-1483-2024 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1483-2024%22&limit=1).

The FDA announced on March 6, 2024, that Microbiologics Inc is recalling KWIK-STIK Rhizopus stolonifer (+) quality control kits due to contamination. The positive control material was contaminated with Aspergillus brasiliensis, which may lead to incorrect test results.

How to tell if you have it

  • Brand: Microbiologics
  • Product: KWIK-STIK Rhizopus stolonifer (+) derived from ATCC 6227b
  • REF 0209P, UDI/DI 20845357003113, Lot Numbers: 209-37-1, 209-37-3, 209-37-4, 209-37-5
  • REF 0209K, UDI/DI 30845357003127, Lot Numbers: 209-37-2

The hazard

The positive control material (Rhizopus stolonifer) within the KWIK-STIK assembly was contaminated with another fungal organism (Aspergillus brasiliensis). As a result, users may observe A. brasiliensis growth when using the control material.

Context

GovInfoNow's records show Microbiologics Inc has 35 other recalls on record since June 6, 2017. Other recent recalls include Microbiologics KWIK-STIK 2-Pack (Dec. 2, 2025), Microbiologics KWIK-STIK Quality Control Kit (Nov. 13, 2025), and Microbiologics LYFO-DISK (June 27, 2025). Other recent recalls of similar products include Medline Convenience Kits (Aug. 11, 2026), Medline Blood Culture Kit (Dec. 24, 2025), and Hologic ThinPrep CytoLyt Solution (Dec. 12, 2025).

How to tell if you have this one

Check for the product name 'KWIK-STIK Rhizopus stolonifer (+) derived from ATCC 6227b'. The kits have model numbers REF 0209P (lot numbers 209-37-1 to 209-37-5) or REF 0209K (lot number 209-37-2).

  • Brand
    Microbiologics
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Check Product Details

    Identify if your KWIK-STIK kit has model REF 0209P with lot numbers 209-37-1 to 209-37-5 or REF 0209K with lot number 209-37-2.

Where it went and when

Where it was distributed

Distributed nationwide and abroad; the notice names 6 states.

Alaska: nationwideAKMaine: nationwideMEVermont: nationwideVTNew Hampshire: nationwideNHWashington: nationwideWAIdaho: nationwideIDMontana: nationwideMTNorth Dakota: nationwideNDMinnesota: nationwideMNIllinois: nationwideILWisconsin: nationwideWIMichigan: named in the noticeMINew York: nationwideNYRhode Island: nationwideRIMassachusetts: nationwideMAOregon: nationwideORNevada: nationwideNVWyoming: nationwideWYSouth Dakota: nationwideSDIowa: nationwideIAIndiana: nationwideINOhio: nationwideOHPennsylvania: nationwidePANew Jersey: nationwideNJConnecticut: nationwideCTCalifornia: named in the noticeCAUtah: nationwideUTColorado: nationwideCONebraska: nationwideNEMissouri: named in the noticeMOKentucky: named in the noticeKYWest Virginia: nationwideWVVirginia: nationwideVAMaryland: nationwideMDDelaware: nationwideDEArizona: nationwideAZNew Mexico: named in the noticeNMKansas: nationwideKSArkansas: nationwideARTennessee: named in the noticeTNNorth Carolina: nationwideNCSouth Carolina: nationwideSCDistrict of Columbia: nationwideDCOklahoma: nationwideOKLouisiana: nationwideLAMississippi: nationwideMSAlabama: nationwideALGeorgia: nationwideGAHawaii: nationwideHITexas: nationwideTXFlorida: nationwideFLPuerto Rico: named in the noticePR

FDA distribution: “Worldwide - US Nationwide distribution in the states of CA, MI, KY, MO, NM, TN, PR, and the countries of France, Columbia, Kazakhstan, Peru, Puerto Rico, South Africa.” States read from FDA’s own wording; “not listed” means the notice does not name it.

Named in the notice: California, Kentucky, Michigan, Missouri, New Mexico, Puerto Rico, Tennessee.

Recall timeline

FDA published this recall 42 days after the firm started it.

1. Firm started the recall: Mar 6, 202412. FDA classified it Class II: Apr 5, 202423. Published in FDA’s Enforcement Report: Apr 17, 202432024
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-1483-2024 (March 6, 2024) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-1483-2024 to see it in context.
  • Plain-language summary written by qwen3:4b on August 19, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA