Skip to content
Lower risk Medical device recall · FDA · Published March 7, 2024

Olympus POWERSEAL 5mm 44cm Sealer: Incomplete Seal Cycle Causes Surgery Delays

Olympus recalls POWERSEAL 5mm 44cm curved jaw sealers due to incomplete seal cycles causing surgery delays. Affected units have model PS-0544CJDA and lot code CA179370.

🇺🇸 FDA recall #Z-1701-2024Medical deviceOlympus ›
Take action

Contact the company

Recalling firmOlympus Corporation of the Americas3500 Corporate Pkwy · Center Valley, PA 18034-8229
✎ GovInfoNow plain-language explainer

Olympus is recalling POWERSEAL surgical sealers that may not seal properly during surgery, causing delays. This could impact surgical timing and patient care. Read more at govinfonow.com/recall/55953-olympus-powerseal-5mm-44cm-sealer-incomplete-seal-cycle-causes-surgery.

Source Summary by GovInfoNow · govinfonow.com/recall/55953-olympus-powerseal-5mm-44cm-sealer-incomplete-seal-cycle-causes-surgery · Updated August 17, 2026 · FDA enforcement record: FDA #Z-1701-2024

Cite this page

“Olympus POWERSEAL 5mm 44cm Sealer: Incomplete Seal Cycle Causes Surgery Delays.” GovInfoNow, updated August 17, 2026. https://govinfonow.com/recall/55953-olympus-powerseal-5mm-44cm-sealer-incomplete-seal-cycle-causes-surgery. Based on FDA recall #Z-1701-2024 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1701-2024%22&limit=1).

The FDA announced on March 7, 2024, that Olympus Corporation of the Americas is recalling its POWERSEAL 5 mm, 44 cm, Curved Jaw Sealer and Divider. The device may fail to deliver energy upon activation, causing a delay in surgery.

How to tell if you have it

  • Brand: Olympus
  • Product: POWERSEAL 5 mm, 44 cm, Curved Jaw Sealer and Divider
  • Model/Catalog Number: PS-0544CJDA
  • UDI-DI: 00821925044579
  • Lot Code: CA179370

The hazard

Upon initial activation of the seal function, the device may produce an Incomplete Seal Cycle tone and display a message on the generator screen without delivering energy. This non-conformance results in a delay in surgery.

Context

GovInfoNow's records show Olympus Corporation of the Americas has issued 133 other recalls since August 18, 2017. Recent brand recalls include the Olympus Tripp-Lite P580-025 DisplayPort Cable (August 27, 2026) and the Olympus M3 Gold Autoclavable Foroblique Telescope 30 (August 24, 2026). Other recent recalls of similar products include Medline convenience kits with Olympus biopsy valves (February 16, 2026) and the Olympus OER-Pro (October 31, 2025).

How to tell if you have this one

Check for model number PS-0544CJDA and lot code CA179370 on the device. These units were sold with the UDI-DI 00821925044579.

  • Brand
    Olympus
  • Category
    Medical device
FDA recall event · 3 products

All 3 products in this recall

FDA lists each product line separately under one recall event. Check your product description, lot or code numbers against each row. Every row shares the hazard and remedy on this page.

  • Classification
    Class II
  • Status
    Ongoing
  • Distribution
    Worldwide distribution - US Nationwide and the countries of AU, CA, HK, KR.
Recall numberProduct, codes & lotsQuantity
Z-1699-2024POWERSEAL 5 mm, 23 cm, Curved Jaw Sealer and Divider, Double Action is indicated for use in general surgery and such surgical specialties as urologic, colorectal, bariatric, vascular, thoracic, and gynecologic. Model/Catalog Number: PS-0523CJDA; UDI-DI: 00821925044531 Lot number: CA182554160 units: (20 units US); (140 units OUS)
Z-1700-2024POWERSEAL 5 mm, 37 cm, Curved Jaw Sealer and Divider, Double Action is indicated for use in general surgery and such surgical specialties as urologic, colorectal, bariatric, vascular, thoracic, and gynecologic. Model/Catalog Number: PS-0537CJDA; UDI-DI: 00821925044555 Lot numbers: CA182543, CA182544, CA182546, CA182548, CA191230850 units: (400 units US);(450 units OUS)
Z-1701-2024POWERSEAL 5 mm, 44 cm, Curved Jaw Sealer and Divider, Double Action is indicated for use in general surgery and such surgical specialties as urologic, colorectal, bariatric, vascular, thoracic, and gynecologic. Model/Catalog Number: PS-0544CJDA; UDI-DI:00821925044579 Lot Code: CA179370150 units: (20 units US); (130 units OUS)
Based on the official FDA remedy

What to do — step by step

  1. Check Device Status

    Verify if the device emits an incomplete seal cycle tone and shows no energy delivery during activation.

Where it went and when

Where it was distributed

Distributed nationwide, and outside the U.S.

Alaska: yesAKMaine: yesMEVermont: yesVTNew Hampshire: yesNHWashington: yesWAIdaho: yesIDMontana: yesMTNorth Dakota: yesNDMinnesota: yesMNIllinois: yesILWisconsin: yesWIMichigan: yesMINew York: yesNYRhode Island: yesRIMassachusetts: yesMAOregon: yesORNevada: yesNVWyoming: yesWYSouth Dakota: yesSDIowa: yesIAIndiana: yesINOhio: yesOHPennsylvania: yesPANew Jersey: yesNJConnecticut: yesCTCalifornia: yesCAUtah: yesUTColorado: yesCONebraska: yesNEMissouri: yesMOKentucky: yesKYWest Virginia: yesWVVirginia: yesVAMaryland: yesMDDelaware: yesDEArizona: yesAZNew Mexico: yesNMKansas: yesKSArkansas: yesARTennessee: yesTNNorth Carolina: yesNCSouth Carolina: yesSCDistrict of Columbia: yesDCOklahoma: yesOKLouisiana: yesLAMississippi: yesMSAlabama: yesALGeorgia: yesGAHawaii: yesHITexas: yesTXFlorida: yesFL

FDA distribution: “Worldwide distribution - US Nationwide and the countries of AU, CA, HK, KR.” States read from FDA’s own wording; “not listed” means the notice does not name it.

Distributed nationwide, and outside the U.S.

Recall timeline

FDA published this recall 2 months after the firm started it.

1. Firm started the recall: Mar 7, 202412. FDA classified it Class II: Apr 30, 202423. Published in FDA’s Enforcement Report: May 8, 202432024
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-1701-2024 (March 7, 2024) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-1701-2024 to see it in context.
  • Plain-language summary written by qwen3:4b on August 17, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA