Skip to content
Lower risk Medical device recall · FDA · Published April 22, 2024

Murata V: Model 2050 Bedside Cardiac Monitor May Show Distorted Vital Signs

Murata Vios recalls Model 2050 bedside cardiac monitors where patient vital signs and ECG waveforms may flicker or flash during setup, affecting accuracy in monitoring.

🇺🇸 FDA recall #Z-1069-2025Medical deviceMurata Vios ›
Take action

Contact the company

Recalling firmMurata Vios, Inc.700 Commerce Dr Ste 190 · Woodbury, MN 55125-9239
✎ GovInfoNow plain-language explainer

Murata Vios is recalling its Model 2050 bedside cardiac monitors because patient vital signs and ECG waveforms can become distorted and flicker during setup. This issue only affects the Central Station Monitor view and not other monitoring areas, so it may not cause immediate harm but could lead to inaccurate readings. Read more at govinfonow.com/recall/55340-murata-v-model-2050-bedside-cardiac-monitor-may-show-distorted.

Source Summary by GovInfoNow · govinfonow.com/recall/55340-murata-v-model-2050-bedside-cardiac-monitor-may-show-distorted · Updated August 18, 2026 · FDA enforcement record: FDA #Z-1069-2025

Cite this page

“Murata V: Model 2050 Bedside Cardiac Monitor May Show Distorted Vital Signs.” GovInfoNow, updated August 18, 2026. https://govinfonow.com/recall/55340-murata-v-model-2050-bedside-cardiac-monitor-may-show-distorted. Based on FDA recall #Z-1069-2025 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1069-2025%22&limit=1).

Murata Vios, Inc. voluntarily recalled the muRata Vios Monitoring System Model 2050 bedside cardiac monitor on April 22, 2024, according to the FDA. The hazard involves distorted patient vital signs and ECG waveforms that flicker or flash rapidly during setup.

How to tell if you have it

  • Brand: Murata Vios
  • Product: muRata Vios Monitoring System Model 2050; bedside cardiac monitor, including cardiotachometer and alarm rate
  • Software version: 1.57.20
  • UDI: +B595BSM20500/$$7V1.57.20/16D20231129Y
  • Quantity: 62 units
  • Distribution: Worldwide, including NY, TX, OH, MD, and India

The hazard

During the set-up and workflow to begin Vios monitoring, patient vital signs and ECG waveforms can become distorted and will flicker or flash very fast. This issue only occurs on the Central Station Monitor (CSM) in the patient tiles view and is not seen at any other point during patient monitoring or on the Bedside Monitor.

Context

GovInfoNow's records show Murata Vios, Inc. has had three other recalls since June 21, 2021. A previous recall of the same Model 2050 monitor occurred on June 21, 2021, citing distorted vital signs during setup. Other recent recalls of similar products include Philips Cardiac Workstation 5000 (July 1, 2026), Philips IntelliVue MX40 (Oct. 31, 2025), and Spacelabs Healthcare Multi-parameter Command Module Model 91496 (Feb. 27, 2026).

How to tell if you have this one

Check for the Model 2050 bedside cardiac monitor with software version 1.57.20. The product includes a cardiotachometer and alarm rate features. The recall applies to units sold with the UDI code +B595BSM20500/$$7V1.57.20/16D20231129Y.

  • Brand
    Murata Vios
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Check Software Version

    Verify your monitor's software version is 1.57.20 by checking the device settings or label.

  2. Contact Manufacturer

    Reach out to Murata Vios directly for guidance on whether your unit is affected and next steps.

Where it went and when

Where it was distributed

Distributed nationwide and abroad; the notice names 4 states.

Alaska: nationwideAKMaine: nationwideMEVermont: nationwideVTNew Hampshire: nationwideNHWashington: nationwideWAIdaho: nationwideIDMontana: nationwideMTNorth Dakota: nationwideNDMinnesota: nationwideMNIllinois: nationwideILWisconsin: nationwideWIMichigan: nationwideMINew York: named in the noticeNYRhode Island: nationwideRIMassachusetts: nationwideMAOregon: nationwideORNevada: nationwideNVWyoming: nationwideWYSouth Dakota: nationwideSDIowa: nationwideIAIndiana: nationwideINOhio: named in the noticeOHPennsylvania: nationwidePANew Jersey: nationwideNJConnecticut: nationwideCTCalifornia: nationwideCAUtah: nationwideUTColorado: nationwideCONebraska: nationwideNEMissouri: nationwideMOKentucky: nationwideKYWest Virginia: nationwideWVVirginia: nationwideVAMaryland: named in the noticeMDDelaware: nationwideDEArizona: nationwideAZNew Mexico: nationwideNMKansas: nationwideKSArkansas: nationwideARTennessee: nationwideTNNorth Carolina: nationwideNCSouth Carolina: nationwideSCDistrict of Columbia: nationwideDCOklahoma: nationwideOKLouisiana: nationwideLAMississippi: nationwideMSAlabama: nationwideALGeorgia: nationwideGAHawaii: nationwideHITexas: named in the noticeTXFlorida: nationwideFL

FDA distribution: “Worldwide - US Nationwide distribution in the states of NY, TX, OH, MD and the country of India.” States read from FDA’s own wording; “not listed” means the notice does not name it.

Named in the notice: Maryland, New York, Ohio, Texas.

Recall timeline

FDA published this recall 9 months after the firm started it.

1. Firm started the recall: Apr 22, 202412. FDA classified it Class II: Jan 30, 202523. Published in FDA’s Enforcement Report: Feb 5, 2025320242025
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-1069-2025 (April 22, 2024) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-1069-2025 to see it in context.
  • Plain-language summary written by qwen3:4b on August 18, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA