Skip to content
Use caution Medical device recall · FDA · Published May 16, 2024

Philips Allura Xper FD10 System with ADN7NT Table Recall: Unintended Table Movement Risk

Philips recalls Allura Xper FD10 systems with ADN7NT patient tables (model 722010, software 8.1.100) due to risk of unintended table movement after software update. Affected units may require repair or replacement.

🇺🇸 FDA recall #Z-1896-2024Medical devicePhilips ›
Take action

Contact the company

Recalling firmPHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Veenpluis 4-6 · Best, Netherlands
✎ GovInfoNow plain-language explainer

Philips is recalling Allura Xper FD10 systems with specific ADN7NT patient tables that may move unexpectedly after a software update. This could lead to patient safety risks if the table shifts during medical procedures. Read more at govinfonow.com/recall/55054-philips-allura-xper-fd10-system-with-adn7nt-table-recall-unintended.

Source Summary by GovInfoNow · govinfonow.com/recall/55054-philips-allura-xper-fd10-system-with-adn7nt-table-recall-unintended · Updated August 19, 2026 · FDA enforcement record: FDA #Z-1896-2024

Cite this page

“Philips Allura Xper FD10 System with ADN7NT Table Recall: Unintended Table Movement Risk.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/55054-philips-allura-xper-fd10-system-with-adn7nt-table-recall-unintended. Based on FDA recall #Z-1896-2024 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1896-2024%22&limit=1).

The FDA announced on May 16, 2024, that Philips Medical Systems Nederland B.V. is recalling the Philips Allura Xper FD10 system with an ADN7NT patient table due to a software issue. The hazard involves unexpected table movements if Software upgrade 8.1.100 is installed.

How to tell if you have it

  • Brand: Philips
  • Product: Philips Allura Xper FD10 system with an ADN7NT patient table
  • Model Number: 722010
  • Software Version Number: 8.1.100
  • Product number: 722010
  • UDI/DI: 00884838059030
  • Serial Numbers: 774, 909, 526

The hazard

When Software upgrade 8.1.100 is installed on an Allura Xper with an AD7NT (non tilt) table, the pivot brake does not engage. This issue may cause unexpected movements of the table.

Context

GovInfoNow's records show PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. has 58 other recalls on record since 2021-05-26. Other recent recalls of similar products include IMRIS ORT300 Operating Room Table Recall (2026-08-06), IMRIS ORT200 Operating Room Table Recall (2026-08-06), and GE Optima XR200amx Mobile Radiography System Recall (2026-06-05).

How to tell if you have this one

Check for model number 722010 and software version 8.1.100 on the Philips Allura Xper FD10 system with an ADN7NT patient table. Serial numbers 774, 909, or 526 are also associated with this recall.

  • Brand
    Philips
  • Category
    Medical device
FDA recall event · 3 products

All 3 products in this recall

FDA lists each product line separately under one recall event. Check your product description, lot or code numbers against each row. Every row shares the hazard and remedy on this page.

  • Classification
    Class II
  • Status
    Ongoing
  • Distribution
    Domestic: OK International: India, Spain, U.A.E.
Recall numberProduct, codes & lotsQuantity
Z-1895-2024Philips Allura Xper FD20 system with an ADN7NT patient table. Model Number: 722012, Software Version Number 8.1.100.; Product number: 722012, UDI/DI: 00884838059054; Serial Number: 2138.1 unit (US)
Z-1896-2024Philips Allura Xper FD10 system with an ADN7NT patient table. Model Number: 722010, Software Version Number 8.1.100.; Product number: 722010, UDI/DI: 00884838059030; Serial Numbers: 774, 909, 526.3 units (OUS)
Z-1897-2024Philips Allura Xper FD20 Biplane system with an ADN7NT patient table. Model Number: 722013, Software Version Number 8.1.100.; Product number: 722013, UDI/DI: 00884838059061; Serial Number: 232.1 unit (OUS)
Based on the official FDA remedy

What to do — step by step

  1. Check Software Version

    Verify the software version on your Philips Allura Xper FD10 system with ADN7NT table. If it is 8.1.100, it is affected.

  2. Contact Philips Support

    Reach out to Philips for guidance on whether your system needs repair or replacement.

Where it went and when

Where it was distributed

Distributed in Oklahoma, and outside the U.S.

Alaska: not listedAKMaine: not listedMEVermont: not listedVTNew Hampshire: not listedNHWashington: not listedWAIdaho: not listedIDMontana: not listedMTNorth Dakota: not listedNDMinnesota: not listedMNIllinois: not listedILWisconsin: not listedWIMichigan: not listedMINew York: not listedNYRhode Island: not listedRIMassachusetts: not listedMAOregon: not listedORNevada: not listedNVWyoming: not listedWYSouth Dakota: not listedSDIowa: not listedIAIndiana: not listedINOhio: not listedOHPennsylvania: not listedPANew Jersey: not listedNJConnecticut: not listedCTCalifornia: not listedCAUtah: not listedUTColorado: not listedCONebraska: not listedNEMissouri: not listedMOKentucky: not listedKYWest Virginia: not listedWVVirginia: not listedVAMaryland: not listedMDDelaware: not listedDEArizona: not listedAZNew Mexico: not listedNMKansas: not listedKSArkansas: not listedARTennessee: not listedTNNorth Carolina: not listedNCSouth Carolina: not listedSCDistrict of Columbia: not listedDCOklahoma: yesOKLouisiana: not listedLAMississippi: not listedMSAlabama: not listedALGeorgia: not listedGAHawaii: not listedHITexas: not listedTXFlorida: not listedFL

FDA distribution: “Domestic: OK International: India, Spain, U.A.E.” States read from FDA’s own wording; “not listed” means the notice does not name it.

States: Oklahoma.

Recall timeline

FDA published this recall 20 days after the firm started it.

1. Firm started the recall: May 16, 202412. FDA classified it Class II: May 24, 202423. Published in FDA’s Enforcement Report: Jun 5, 202432024
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-1896-2024 (May 16, 2024) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-1896-2024 to see it in context.
  • Plain-language summary written by qwen3:4b on August 19, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA