Skip to content
Use caution Medical device recall · FDA · Published May 22, 2024

Beckman Coulter DxI 9000 Access Analyzer Software Risk: Sample Timeout Errors

Beckman Coulter recalls DxI 9000 Access Immunoassay Analyzers with software versions 1.17.0 or below due to sample timeout errors that may delay patient treatment. Affected units have specific catalog numbers and UDI-DI codes.

🇺🇸 FDA recall #Z-2735-2024Medical deviceBeckman Coulter ›
Take action

Contact the company

Recalling firmBeckman Coulter, Inc.1000 Lake Hazeltine Dr · Chaska, MN 55318-1037
✎ GovInfoNow plain-language explainer

Beckman Coulter is recalling DxI 9000 Access Immunoassay Analyzers with software versions 1.17.0 or older. This issue can cause sample tests to be flagged as timeout errors, potentially delaying patient treatment if the analyzer runs out of sample capacity. Read more at govinfonow.com/recall/55034-beckman-coulter-dxi-9000-access-analyzer-software-risk-sample-timeout.

Source Summary by GovInfoNow · govinfonow.com/recall/55034-beckman-coulter-dxi-9000-access-analyzer-software-risk-sample-timeout · Updated August 19, 2026 · FDA enforcement record: FDA #Z-2735-2024

Cite this page

“Beckman Coulter DxI 9000 Access Analyzer Software Risk: Sample Timeout Errors.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/55034-beckman-coulter-dxi-9000-access-analyzer-software-risk-sample-timeout. Based on FDA recall #Z-2735-2024 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2735-2024%22&limit=1).

The FDA announced on May 22, 2024, that Beckman Coulter is recalling DxI 9000 Access Immunoassay Analyzers due to a software defect that may delay test results and patient treatment. The agency issued recall number Z-2735-2024 for the Class II voluntary recall.

How to tell if you have it

  • Brand: Beckman Coulter
  • Product: DxI 9000 Access Immunoassay Analyzer
  • Catalog #: C11137
  • Software versions: 1.17.0 or below
  • UDI-DI: 15099590732103

The hazard

According to the FDA, when connected to a Laboratory Automation System, analyzers with software versions 1.17.0 and below may fail to signal that the sample wheel is full. This can cause test results to be flagged with a sample wheel timeout error and delay reporting, potentially leading to delayed patient treatment.

Context

GovInfoNow's records show Beckman Coulter has issued 132 other recalls since June 20, 2017. Recent similar product recalls include the AU480 Chemistry Analyzer recall on August 3, 2026, and the AU5800 Chemistry Analyzer recall on January 30, 2026.

How to tell if you have this one

Check for the catalog number C11137 and software versions 1.17.0 or below. The device also has a specific UDI-DI code 15099590732103.

  • Brand
    Beckman Coulter
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Update Software

    Install the latest software version to fix the timeout error issue.

Where it went and when

Where it was distributed

Distributed nationwide and abroad; the notice names 12 states.

Alaska: nationwideAKMaine: named in the noticeMEVermont: nationwideVTNew Hampshire: named in the noticeNHWashington: named in the noticeWAIdaho: nationwideIDMontana: nationwideMTNorth Dakota: nationwideNDMinnesota: nationwideMNIllinois: nationwideILWisconsin: nationwideWIMichigan: nationwideMINew York: nationwideNYRhode Island: nationwideRIMassachusetts: nationwideMAOregon: nationwideORNevada: nationwideNVWyoming: nationwideWYSouth Dakota: nationwideSDIowa: nationwideIAIndiana: named in the noticeINOhio: nationwideOHPennsylvania: named in the noticePANew Jersey: named in the noticeNJConnecticut: nationwideCTCalifornia: named in the noticeCAUtah: nationwideUTColorado: nationwideCONebraska: nationwideNEMissouri: nationwideMOKentucky: nationwideKYWest Virginia: nationwideWVVirginia: nationwideVAMaryland: nationwideMDDelaware: nationwideDEArizona: nationwideAZNew Mexico: nationwideNMKansas: named in the noticeKSArkansas: nationwideARTennessee: nationwideTNNorth Carolina: nationwideNCSouth Carolina: nationwideSCDistrict of Columbia: nationwideDCOklahoma: named in the noticeOKLouisiana: nationwideLAMississippi: nationwideMSAlabama: nationwideALGeorgia: named in the noticeGAHawaii: nationwideHITexas: named in the noticeTXFlorida: named in the noticeFL

FDA distribution: “Worldwide - US Nationwide distribution in the states of CA, FL, GA, IN, KS, ME, NH, NJ, OK, PA, TX, and WA. The countries of Australia, Austria, Bahrain, Belgium, Brazil, Croatia, Czechia, Egypt, France, Germany, Hungary, Ireland, Israel,…” States read from FDA’s own wording; “not listed” means the notice does not name it.

Named in the notice: California, Florida, Georgia, Indiana, Kansas, Maine, New Hampshire, New Jersey, Oklahoma, Pennsylvania, Texas, Washington.

Recall timeline

FDA published this recall 3 months after the firm started it.

1. Firm started the recall: May 22, 202412. FDA classified it Class II: Aug 27, 202423. Published in FDA’s Enforcement Report: Sep 4, 202432024
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-2735-2024 (May 22, 2024) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-2735-2024 to see it in context.
  • Plain-language summary written by qwen3:4b on August 19, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA