Siemens recalls AXIOM MULTIX MT X-ray tables due to short circuit in power supply causing fire risk. Affected models have specific serial numbers. Contact Siemens for remedy.
Short circuit in the power supply unit (PSU) of the Touch Display may lead to overheating and fire
Siemens is recalling AXIOM MULTIX MT X-ray tables because a short circuit in the power supply can cause overheating and fire. This affects specific models sold with Siemens Multix PRO systems.
The power supply unit (PSU) in the touch display can short circuit, leading to overheating and potentially causing a fire. This hazard occurs during normal operation of the X-ray table.
Healthcare facilities using Siemens AXIOM MULTIX MT X-ray tables with specific serial numbers are affected. Patients and staff at these facilities are at risk of fire.
Check the serial numbers on the table: 1010, 1011, 1015, 1105, 1117, 1124, or 1126. The table is used with Siemens Multix PRO and TOP Systems radiographic devices.
Immediately stop using the affected AXIOM MULTIX MT X-ray table as it poses a fire risk.
A short circuit in the power supply unit (PSU) of the touch display can lead to overheating and fire.
Serial numbers 1010, 1011, 1015, 1105, 1117, 1124, and 1126.
Check if your table has the serial numbers listed: 1010, 1011, 1015, 1105, 1117, 1124, or 1126. It is used with Siemens Multix PRO and TOP Systems radiographic devices.
We’ve drafted a message you can send Siemens to request your refund or repair — edit it as you like.
Subject: Recall Z-2181-2024 — Siemens Siemens Hello, I own a Siemens that is covered by recall Z-2181-2024 from Siemens. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.