Siemens recalls AXIOM MULTIX MP X-ray tables due to short circuit in power supply causing fire risk. Affected models have specific serial numbers. Contact Siemens for return or repair.
Short circuit in the power supply unit (PSU) of the Touch Display may lead to overheating and fire
Siemens is recalling AXIOM MULTIX MP X-ray tables that could overheat and catch fire due to a power supply short circuit. This hazard affects medical facilities using these tables with Siemens Multix PRO systems.
A short circuit in the power supply unit can cause overheating, which may lead to fire. This hazard occurs when the touch display is powered on and the short circuit is active.
Medical facilities using Siemens AXIOM MULTIX MP X-ray tables with Multix PRO systems are most at risk. Patients and staff could be affected if the table malfunctions during use.
Check for serial numbers: 1022, 1074, 1118, 1119, 1120, 1121, 1123, 1126, 1130, 1131, 1134, 1138. The table is used with Siemens Multix PRO and TOP Systems radiographic devices.
Look for the specific serial numbers listed: 1022, 1074, 1118, 1119, 1120, 1121, 1123, 1126, 1130, 1131, 1134, 1138.
The power supply unit in the touch display may short circuit, causing overheating and fire risk.
The recall record does not specify a remedy, so contact Siemens directly for guidance.
The recall indicates a fire risk, so it is not safe to use until the issue is resolved.
We’ve drafted a message you can send Siemens to request your refund or repair — edit it as you like.
Subject: Recall Z-2182-2024 — Siemens Siemens Hello, I own a Siemens that is covered by recall Z-2182-2024 from Siemens. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.