Siemens recalls Multix TOP I PRO X-ray tables with serial numbers 1004, 1006, 1011, 1014 due to fire risk from power supply short circuits. Affected units may overheat and cause fire.
Short circuit in the power supply unit (PSU) of the Touch Display may lead to overheating and fire
Siemens is recalling specific Multix TOP I PRO X-ray tables that could overheat and start a fire due to a power supply short circuit. This recall affects hospitals and clinics using these tables with Siemens Multix PRO and TOP systems.
The power supply unit in the Touch Display can short circuit, causing overheating and potentially starting a fire. This hazard occurs during normal use when the table is powered on.
Hospitals and clinics using the affected Multix TOP I PRO X-ray tables with Siemens Multix PRO and TOP systems are most at risk. The recall applies to units with specific serial numbers.
Check the serial number on the table. Affected units have serial numbers 1004, 1006, 1011, or 1014. The table is used with Siemens Multix PRO and TOP radiographic systems.
Immediately stop using the affected Multix TOP I PRO X-ray table as it poses a fire risk.
The recalled model is the Siemens Multix TOP I PRO Radiographic X-ray Table.
Serial numbers 1004, 1006, 1011, and 1014 are affected.
Check the serial number on the table; affected units have serial numbers 1004, 1006, 1011, or 1014.
We’ve drafted a message you can send Siemens to request your refund or repair — edit it as you like.
Subject: Recall Z-2186-2024 — Siemens Siemens Hello, I own a Siemens that is covered by recall Z-2186-2024 from Siemens. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.