Siemens recalls Multix TOP I PRO X-ray tables with potential fire risk due to power supply short circuit. Affected units have specific serial numbers; contact Siemens for remedy.
Short circuit in the power supply unit (PSU) of the Touch Display may lead to overheating and fire
Siemens is recalling Multix TOP I PRO X-ray tables that may overheat and catch fire due to a power supply short circuit. This could cause serious injury or death if the table is used without proper safety measures.
A short circuit in the power supply unit can cause overheating, which may lead to fire. This hazard occurs when the table is in normal use with electrical connections, potentially causing burns or fire if the short circuit is not fixed.
Healthcare facilities using Siemens Multix TOP I PRO X-ray tables with serial numbers 1523, 1585, or 1897 are affected. Staff and patients using these tables are at risk of fire or injury.
Check for serial numbers 1523, 1:585, or 1897 on the Multix TOP I PRO X-ray table. The table is used with Siemens Multix PRO and TOP Systems radiographic devices.
Immediately stop using the affected Multix TOP I PRO X-ray table to prevent fire risk.
The power supply unit may have a short circuit that causes overheating and fire, posing a serious risk to users.
Contact Siemens immediately to stop using it and follow their instructions for remedy.
Yes, affected units have serial numbers 1523, 1585, or 1897.
We’ve drafted a message you can send Siemens to request your refund or repair — edit it as you like.
Subject: Recall Z-2187-2024 — Siemens Siemens Hello, I own a Siemens that is covered by recall Z-2187-2024 from Siemens. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.