The FDA announced on June 3, 2024, that Qiagen Sciences LLC is recalling QIAcube Connect MDx instruments because the heater shaker module may fail to heat properly. This defect could cause delayed or erroneous test results for users of specific IVD protocols. Owners should contact the company for information.
How to tell if you have it
- Brand: Qiagen
- Product: QIAcube Connect MDx - IVD Instrument designed to perform automated isolation and purification of nucleic acids in
- Catalog Number: 900370
- Software Version: 1.0.1
- GTIN: 04053228039129
- Serial Numbers: SN 30000-30506
- Quantity: 419 units
The hazard
The heater shaker module does not perform heating if the temperature is set below 40°C for static incubation steps when running IVD protocol QIAamp DSP DNA Mini Kit (cat. no. 61304). This malfunction could potentially cause delayed or erroneous results depending on the downstream assay, according to the FDA.
Context
GovInfoNow's records show Qiagen Sciences LLC has 16 other recalls on record since July 23, 2018. Recent similar recalls include the QIAstat-Dx Respiratory Panel Plus recall on March 12, 2025, and the EZ2 Connect MDx-IVD Recall on March 4, 2024. Other recent recalls of similar products include the Covidien Mon-a-Therm Temperature Probe Recall (April 23, 2026) and Medline Convenience Kits Recalls (May 28, 2026).