The FDA announced on June 24, 2024, that Covidien is recalling Puritan Bennett 500 Series Ventilators due to a potential health hazard from volatile organic compounds released after extended use. The agency stated that cumulative use exceeding 14 years may pose harm to health.
How to tell if you have it
- Puritan Bennett 560 Ventilator/4096600
- Puritan Bennett PB560 VENTILATOR EU-DIV/4096600-01
- Puritan Bennett 560 Ventilator JAPAN/4096600-02
- Puritan Bennett 560 Ventilator APAC/4096600-03
- Puritan Bennett 560 Ventilator LATAM/4096600-04
- Puritan Bennett 560 Ventilator EMEA/4096600-05
- Puritan Bennett 560 Ventilator CAN ANZ/4096600-06
- Puritan Bennett 540 Ventilator USA/4097100
- Puritan Bennett 520 Ventilator/4098300
- Puritan Bennett 520 Ventilator EU-DIV/4098300-01
- Puritan Bennett 520 Ventilator Japan/4098300-02
- Puritan Bennett 520 Ventilator APAC/4098300-03
- Puritan Bennett 520 Ventilator EMEA/4098300-05
- Puritan Bennett 520 Ventilator CAN/ANZ/4098300-06
- DL4096600 DL PB560 VENTILATOR/DL4096600
- DL4096600-06 RFB PB560 VENT CAN ANZ/DL4096600-06
- REFURBISHED PB540 VENTILATOR/DL4097100
- Puritan Bennett 540 Ventilator Refurbished/DS4097100
- All device serial numbers used before distribution of PB500 Series Ventilator User Manual and Clinician's Manual addendum, PT00196729 A, 2024-06-11
The hazard
According to the FDA, using these devices for more than 14 years of cumulative duration may pose harm to health due to a volatile organic compound (VOC), 2-propanol, 1,3-dichloro released from some components in the ventilator gas pathway. The record does not report specific incidents or injuries.
Context
GovInfoNow's records show Covidien has 14 other recalls on record since 2021-03-05. Other recent recalls of similar products include Covidien Bedside SpO2 Monitor Recall (2025-06-23), Covidien Newport HT70 Ventilator Recall (2025-05-21), and Covidien Shiley Tracheostomy Tube Recall (2025-02-26). Other recent recalls of similar products include Ventec V+Pro Ventilator Recall (2026-03-25), Baxter Blue Ventilator Adapter Module Recall (2026-02-18), and Burlington Medical Blockers Recall (2026-02-11).