GE Healthcare Centricity PACS Recall (Class II) - August 2024
Recall of 2,138 units of GE Healthcare Centricity PACS devices distributed worldwide. Class II recall issued on August 5, 2024.
Lower-risk recall
GE HealthCare has become aware of an issue where image storage is not possible when Centricity PACS reaches the maximum number of unique image identifiers (Ckeys). If this occurs, it can result in acquisition of a partial study.
Contact the company
What happened, in plain English
GE Healthcare has issued a Class II recall for 2,138 units of its Centricity PACS (Picture Archiving and Communication System) devices due to worldwide distribution concerns. The recall was initiated on August 5, 2024. Read more at govinfonow.com/recall/54319-ge-healthcare-centricity-pacs-v2-v7-recall-image-storage-failure.
Source Summary by GovInfoNow · govinfonow.com/recall/54319-ge-healthcare-centricity-pacs-v2-v7-recall-image-storage-failure · Updated October 13, 2026 · FDA enforcement record: FDA #Z-3214-2024
Cite this page
“GE Healthcare Centricity PACS Recall (Class II) - August 2024.” GovInfoNow, updated October 13, 2026. https://govinfonow.com/recall/54319-ge-healthcare-centricity-pacs-v2-v7-recall-image-storage-failure. Based on FDA recall #Z-3214-2024 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-3214-2024%22&limit=1).
Why it’s dangerous
The recall is classified as Class II, indicating that use of or exposure to the defective product might cause temporary health problems or pose a slight threat of a serious nature. Specific technical reasons for the defect are not detailed in the provided text, but Class II recalls typically involve potential safety risks that require corrective action.
Who’s affected
Healthcare facilities and institutions worldwide that have installed or are using GE Healthcare Centricity PACS devices with the specific serial numbers listed in the recall notice.
How to tell if you have this one
Check if your facility's Centricity PACS devices have any of the following serial numbers or identifiers: - YP5025-Centricity PACS-02170316, JA0842, JL1419, etc. - YP4000-Centricity PACS-02321044, JL1701, JD1247, etc. - YP5020-Centricity PACS-02097541, JDS003, YP5013-Centricity PACS-18045997, etc. - YP5037-Centricity PACS-15349241, JDS058, JL1412, etc. - YP5022-Centricity PACS-02285739, JDS033, JL1460, etc. - YP5049-Centricity PACS-20978976, YP5030-Centricity PACS-02274229, JDS032, etc. - YP4011-Centricity PACS-20584824, YP1017-Centricity PACS-02344060, JL1399, etc. - YP5009-JP-532345-Centricity PACS-20574025, YP4002-Centricity PACS-14539502, JL1168, etc. - YP5042-Centricity PACS-20101281, YP5031-Centricity PACS-02332510, JDS042, etc. - YP1012-Centricity PACS-20584811, JL1152, JL1721, JDS062, etc. - YP4001-JP-407999-Centricity PACS-19736997, JL1730, JDS111, etc. - YP5012-JP-100014-Centricity PACS-20574034, JL1407, JD1065, etc. - YP1014-Centricity PACS-01978327, JL1747, JD1099, etc. - YP1013-Centricity PACS-20584812, JL1102, JL1413, JD1148, etc. - YP5002-JP-341108-Centricity PACS-20568516, YP4006-Centricity PACS-20584822, JL1740, etc. - YP5048-Centricity PACS-20644650, JDS146, JL1736, etc. - YP5034-Centricity PACS-14194563, JDS047, YP5039-Centricity PACS-19629898, etc. - YP5027-Centricity PACS-02255460, JDS019, JDS020, etc. - YP5032-Centricity PACS-02360541, JDS006, YP5014-JP-280363-Centricity PACS-20574036, etc. - YP1020-Centricity PACS-20584818, JL1753, YP5038-Centricity PACS-16450968, etc. - YP5007-JP-280203-Centricity PACS-20574023, YP5004-JP-080362-Centricity PACS-20568519, JL1754, etc. - YP4012-JP-080218-Centricity PACS-20579603, JL1743, JD1091, etc. - YP1011-JP-081007-Centricity PACS-20579601, JD1160, JD1167, etc. - YP5003-JP-080145-Centricity PACS-20568518, JL1751, JD1051, etc. - JL1601-Centricity PACS-20584807, YP5024-Centricity PACS-02179873, JDS010, etc. - YP5010-JP-143756-Centricity PACS-20574031, JL1392, JD0999, etc. - YP5023-Centricity PACS-02256894, JAS006, JDS026, etc. - YP5021-JP-240130-Centricity PACS-20574046, YP5001-JP-040967-Centricity PACS-20568514, JL1744, etc. - YP5006-JP-040008-Centricity PACS-20568522, JA0741, JD1161, etc. - YP5035-Centricity PACS-15111123, JL1173, JL1390, etc. - YP5019-Centricity PACS-02042638, JL1749, JD1154, etc. - YP5000-Centricity PACS-02277426, YP5036-Centricity PACS-15261448, JDS056, etc. - YP4004-Centricity PACS-20584821, YP5050-Centricity PACS-21180376, JL1742, etc. - YP4005-JP-110249-Centricity PACS-20535391, JL1405, JD1057, etc. - YP5026-Centricity PACS-02253241, JDS017, JDS018, etc. - YP5033-Centricity PACS-02399305, JAS011, JDS044, etc. - YP4003-Centricity PACS-02276197, JDS004, YP5015-JP-221077-Centricity PACS-20574038, etc. - YP5018-Centricity PACS-02013203, JL1722, JAS018, etc. - YP1005-Centricity PACS-17452182, JL1391, JD0990, etc. - YP1004-Centricity PACS-20584809, JL1153, JD0784, etc. - YP5044-Centricity PACS-20484083, JAS027, YP5005-JP-134658-Centricity PACS-20568520, etc. - YP5028-Centricity PACS-02248290, JL1174, JD0839, etc. - YP5011-Centricity PACS-02178701, JDS124, YP5046-Centricity PACS-20578187, etc. - YP5016-JP-130422-Centricity PACS-20574042, JL1169, JD1267, etc. - YP5017-JP-160102-Centricity PACS-20574044, JD1050, JD1062, etc. - YP5045-Centricity PACS-20509113, JL1430, JD1071, etc. - YP1006-Centricity PACS-02309861, JL1735, JL1717, etc. Additionally, look for devices associated with the following hospital/institution identifiers: - KE2701CP01, KE0802CP01, KE0402CP01, etc. - 0850550196, 0850550301, 0850550115, etc. - 852004_RAD-Centricity PACS-20542654, KW1190PC03, KW1011PC04, etc. - LB1160SV01, LB1439PC02, LB1462PC01, etc. - MMC2, AMPANG6, IPPT4, SERDANG1, UTAR-CPACS, HSB-CPACS, etc. - HCIT4579183PACS, HCIT4665855PACS, HCIT4376494PACS, etc. - RU6265ID05, RF1647ID05, tobecreated, RU3143ID01, etc. - SA2706PP05, SA2247PP02, SA1246PP02, etc. - NNI-SG-CPACS, KTPH8, ASIAMEDIC02, RADLINK10, AH-CPACS, etc. - 4547195 - 6966037 - 3227712-Centricity PACS-24283964, NTFGH3, NUH-PACS, etc. - ZA1352PP05, ZA2241PAC3, ZA2895PP02, etc. - FPG10802, FPG10902, EPP05803, FEG85600, FEG85608, etc. - MPX06502, MPX06503, MPX23600, MPX91100, MPX43176, etc. - NE465120, NE670325, NE673137, NE300609, NE219643, etc. - FPD10805, FPD10806, FPD10807, FE816139, FE816140, etc. - DPP03300, DPP03301, DPP03302, DPP03303, DPP03305, etc. - FEG51800, FEF70522, FEF70523, YE151130, YE162A34, etc. - GPX45803, GPX45903, MEX70164, MEX67400, MEX25523, etc. - GEX18404, GEQ20315, GEX00300, GEC21836, GE448033, etc. - DPN08303, DEN27229, DEL26465, DEL26479, DE993820, etc. - NPV73804, NE203141, NE740521, NE216225, NE487826, etc. - DPD01000, DE539281, DE539290, DPD33502, FPD36800, etc. - 529010CS02, 150010PA01, 529020PC03, 437918CS01, 437918SV03, etc. - CGMHKEL20541717, 0853550429, 0853550209, 0853550142, etc. - HCIT4783733PACS, P06127PAC2, 06002PAC37, 06172PAC01, etc. - AE0002PP02, AE1802PP01, AE1564PP03, AE1761PAC2, 81564SUN01, etc. - RRJ0500038, 81001VIR04, RWJ0100069, RJN7100049, 80327ISU01, etc. - DDC01VM003, RFSDAVM014, RW60200064, 85235DBS01, RVV0100002, etc. - RPA0200134, RWF0300076, RWF0300084, RVV0900052, RPA1700007, etc. - PRDC0VM045, RNQ51VM014, 80257HOS01, To be Provided, To be Provided, etc. - NNS01HOS01, NNS01TEST1, 85252LOA01, RX1LBTEST1, RK5BCVM002, etc. - BacLieuGH-CPACS, VINMECDN07, VINMECHL07, THANHNHAN01, VINMECHAIPHONG07, etc. - VinUni-CPACS, CHORAY01, BINHCHANHGH-CPACS, VINMECTANCANG-CPACS, etc.
- Brand
GE Healthcare- Category
Medical device
What to do — step by step
,
Identify all Centricity PACS devices in your facility and check their serial numbers against the list provided in the recall notice.
,
If any of your devices match the recalled serial numbers, immediately contact GE Healthcare customer support or the local distributor for further instructions.
,
Do not remove or disable the device from service unless explicitly instructed by GE Healthcare. The recall may involve software updates, hardware replacements, or other corrective actions that can be performed while the device remains in use.
,
Monitor any communications from GE Healthcare regarding the recall process, including timelines for corrective actions and potential impacts on patient care.
,
Ensure that all relevant staff members are aware of the recall and follow any interim safety guidelines provided by GE Healthcare or local health authorities.
How this recall unfolded
Recall timeline
FDA published this recall 58 days after the firm started it.
- Firm started the recall
- FDA classified it Class II
- Published in FDA’s Enforcement Report
Dates from the agency record only; steps without a date are left out.
Questions people ask
What is the Centricity PACS?
Centricity PACS (Picture Archiving and Communication System) is a medical imaging software solution developed by GE Healthcare used for storing, retrieving, managing, and distributing medical images.
Similar recalls
Latest medical device recalls
Keep exploring
Where this comes from
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.
- Official source: FDA Recall #Z-3214-2024 (August 5, 2024) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-3214-2024 to see it in context.
- Plain-language summary written by qwen3.6:35b-a3b on October 13, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
- GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.