Skip to content
Lower risk Medical device recall · FDA · Published August 6, 2024

Philips Azurion 7M20 Systems with FlexArm May Have Unreliable Movement - Recall

Philips recalls Azurion 7M20 systems with FlexArm ceiling-mounted systems due to potential inconsistent motorized movement from grease leakage causing lubrication issues.

🇺🇸 FDA recall #Z-2953-2024Medical devicePhilips ›
Take action

Contact the company

Recalling firmPHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Veenpluis 4-6 · Best, Netherlands
✎ GovInfoNow plain-language explainer

Philips is recalling Azurion 7M20 systems with FlexArm ceiling-mounted systems because the motorized movement may become inconsistent and eventually stop working. This could affect medical procedures requiring precise positioning. Read more at govinfonow.com/recall/54307-philips-azurion-7m20-systems-with-flexarm-may-have-unreliable-movement.

Source Summary by GovInfoNow · govinfonow.com/recall/54307-philips-azurion-7m20-systems-with-flexarm-may-have-unreliable-movement · Updated August 19, 2026 · FDA enforcement record: FDA #Z-2953-2024

Cite this page

“Philips Azurion 7M20 Systems with FlexArm May Have Unreliable Movement - Recall.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/54307-philips-azurion-7m20-systems-with-flexarm-may-have-unreliable-movement. Based on FDA recall #Z-2953-2024 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2953-2024%22&limit=1).

The FDA announced on Aug. 6, 2024, that Philips Medical Systems Nederland B.V. is recalling 927 Philips Azurion 7M20 systems with FlexArm ceiling-mounted systems due to a motorized movement hazard. The agency classified the recall as Class II.

How to tell if you have it

  • Philips Azurion 7M20 systems with FlexArm ceiling-mounted system, Interventional Fluoroscopic X-Ray System
  • Model Numbers: 722079 and 722224
  • UPCs: 00884838085268 and 00884838099258

The hazard

The motorized longitudinal movement of the FlexArm stand may be inconsistent or become unavailable due to grease leakage from bearings and excessive anti-corrosion oil on the ceiling rail. This reduces friction between the rails and the friction wheel.

Context

GovInfoNow's records show Philips Medical Systems Nederland B.V. has had 58 other recalls since May 26, 2021. Recent Philips recalls include software issues affecting Azurion R2.2.10 systems reported on July 16, 2026, and a firmware issue leading to loss of geometry movements on June 29, 2026. Other recent recalls of similar products include Canon Alphenix INFX-8000C X-ray System Recall (Dec. 19, 2025), Canon Alphenix INFX-8000V X-ray System Recall (Dec. 19, 2025), and Canon Alphenix INFX-8000H X-ray System Recall (Dec. 19, 2025).

How to tell if you have this one

Check for model numbers 722079 or 722224 on Philips Azurion 7M20 systems with FlexArm ceiling-mounted systems. The serial numbers listed in the recall notice indicate affected units.

  • Brand
    Philips
  • Product
    Philips Azurion 7M20 systems with FlexArm ceiling-mounted system, Interventional Fluoroscopic X-Ray System, Model
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Check Model Numbers

    Identify if your system has model numbers 722079 or 722224.

Where it went and when

Where it was distributed

Distributed nationwide, and outside the U.S.

Alaska: yesAKMaine: yesMEVermont: yesVTNew Hampshire: yesNHWashington: yesWAIdaho: yesIDMontana: yesMTNorth Dakota: yesNDMinnesota: yesMNIllinois: yesILWisconsin: yesWIMichigan: yesMINew York: yesNYRhode Island: yesRIMassachusetts: yesMAOregon: yesORNevada: yesNVWyoming: yesWYSouth Dakota: yesSDIowa: yesIAIndiana: yesINOhio: yesOHPennsylvania: yesPANew Jersey: yesNJConnecticut: yesCTCalifornia: yesCAUtah: yesUTColorado: yesCONebraska: yesNEMissouri: yesMOKentucky: yesKYWest Virginia: yesWVVirginia: yesVAMaryland: yesMDDelaware: yesDEArizona: yesAZNew Mexico: yesNMKansas: yesKSArkansas: yesARTennessee: yesTNNorth Carolina: yesNCSouth Carolina: yesSCDistrict of Columbia: yesDCOklahoma: yesOKLouisiana: yesLAMississippi: yesMSAlabama: yesALGeorgia: yesGAHawaii: yesHITexas: yesTXFlorida: yesFL

FDA distribution: “Worldwide - US Nationwide distribution.” States read from FDA’s own wording; “not listed” means the notice does not name it.

Distributed nationwide, and outside the U.S.

Recall timeline

FDA published this recall 36 days after the firm started it.

1. Firm started the recall: Aug 6, 202412. FDA classified it Class II: Sep 4, 202423. Published in FDA’s Enforcement Report: Sep 11, 202432024
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-2953-2024 (August 6, 2024) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-2953-2024 to see it in context.
  • Plain-language summary written by qwen3:4b on August 19, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA