B Braun Medical Inc. voluntarily recalled the Infusomat Space Infusion System/Large Volume Pump on Aug. 7, 2024, according to the FDA. The pumps may trigger false occlusion alarms due to electrostatic charge buildup.
How to tell if you have it
- Brand: B Braun
- Product: Infusomat Space Infusion System/ Large Volume Pump
- Product Code: 8713050U
- UDI-DI (GUDID): 04046963716752
- Serial Numbers: 52226, 107529, 123259, 123815, 123843, 139003, 147678, 147721, 147725, 239018, 239165, 239231, 239279, 339337
- Quantity: 14 units
- Distribution: Sold in AL, CA, GA, IL, MI, and TX
The hazard
The pumps have the potential for the upstream occlusion sensor to cease proper function due to buildup of electrostatic charge during operation. This issue triggers false downstream and upstream occlusion alarms. The recall extends a previous recall, RES 92978.
Context
GovInfoNow's records show B Braun Medical Inc. has issued 66 other recalls since Oct. 15, 2013. Recent recalls include a Particulate Matter Hazard for 100mL Sodium Chloride Injection on Aug. 19, 2026, and Lactated Ringer's Injection on Aug. 12, 2026. Other recent recalls of similar products include Fresenius Kabi Ivenix Infusion Pump Recall (Nov. 3, 2025), Spectrum IQ Infusion Pump Recall (Jan. 13, 2026), and Fresenius Kabi Ivenix Infusion System Software Recall (May 6, 2026).