Siemens Atellica Microalbumin Test Kits May Give False Low Results

Siemens Healthcare recalls Atellica CH Microalbumin_2 kits for false low microalbumin readings due to Prozone Effect. Affected units have specific lot numbers and were sold with ADVIA Chemistry systems.

🇺🇸 FDA recall #Z-3169-2024Medical deviceSiemens Healthcare ›
Take action

Contact the company

Recalling firmSiemens Healthcare Diagnostics, Inc.511 Benedict Ave · Tarrytown, NY 10591-5005
✎ GovInfoNow plain-language explainer

Siemens Healthcare is recalling Atellica CH Microalbumin_2 test kits because they may give false low results for microalbumin in urine. This could lead to incorrect health assessments if used with ADVIA Chemistry systems. Read more at govinfonow.com/recall/54227-siemens-atellica-microalbumin-test-kits-may-give-false-low-results.

Source Summary by GovInfoNow · govinfonow.com/recall/54227-siemens-atellica-microalbumin-test-kits-may-give-false-low-results · Updated August 19, 2026 · FDA enforcement record: FDA #Z-3169-2024

Cite this page

“Siemens Atellica Microalbumin Test Kits May Give False Low Results.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/54227-siemens-atellica-microalbumin-test-kits-may-give-false-low-results. Based on FDA recall #Z-3169-2024 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-3169-2024%22&limit=1).

The FDA announced on August 15, 2024, that Siemens Healthcare Diagnostics, Inc. is recalling Atellica CH Microalbumin_2 (¿ALB_2) reagent kits because they may yield erroneously depressed microalbumin results. The recall affects approximately 18,483 units distributed worldwide.

How to tell if you have it

  • Brand: Siemens Healthcare
  • Product: Atellica CH Microalbumin_2 (¿ALB_2)
  • Lot Numbers: 232033, 232128, 232137, 232146, 232147, 242149, 242150, 242365
  • UDI-DI: 00630414596310

The hazard

The reagent kits do not meet the High-Dose Hook Effect claim or Prozone Effect claim stated in the Instructions for Use. This failure may result in erroneously depressed microalbumin patient results when used on Atellica CH, Atellica CI Analyzers, ADVIA 1800 Chemistry Systems, ADVIA 2400 Chemistry Systems, and ADVIA Chemistry XPT Systems.

Context

GovInfoNow's records show Siemens Healthcare Diagnostics, Inc. has 150 other recalls on record since June 7, 2017. Other recent recalls of similar products include Siemens Atellica CH RCRP Test (August 17, 2026), Siemens Atellica CH Urine Albumin Recall (July 6, 2026), and Beckman Coulter Emit 2000 Quinidine Assay Recall (July 13, 2026). Other recent recalls of similar products include Magellan LeadCare II Blood Lead Test Kits, Magellan LeadCare Ultra Blood Lead Test Kit, and Magellan LeadCare Plus Blood Lead Test Kit (all March 13, 2026).

How to tell if you have this one

Check the lot number on the kit packaging. Affected kits have lot numbers 232033, 232128, 232137, 232146, 232147, 242149, 242150, or 242365. They were sold with ADVIA Chemistry systems and Atellica analyzers.

  • Brand
    Siemens Healthcare
  • Category
    Medical device
FDA recall event · 2 products

All 2 products in this recall

FDA lists each product line separately under one recall event. Check your product description, lot or code numbers against each row. Every row shares the hazard and remedy on this page.

  • Classification
    Class II
  • Status
    Ongoing
  • Distribution
    Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belgium, Bosnia Herzeg, Brazil, Canada, Chile, Colombia, Croatia, Cura¿ao, St. Eus, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Hong¿Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Latvia, Lithuania, Malaysia, Maldives, Mexico, Morocco, Netherlands, New¿Zealand, Norway, Oman, P.R.¿China, Pakistan, Paraguay, Peru, Poland, Portugal, Qatar, Republic¿Korea, Romania, Russian¿Fed., Saudi¿Arabia, Serbia, Singapore, Slovakia, Slovenia, South¿Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, U.A.E, United¿Kingdom, Uruguay, Vietnam, Zaire. ¿¿¿¿¿¿
Recall numberProduct, codes & lotsQuantity
Z-3169-2024Atellica CH Microalbumin_2 (¿ALB_2)-For in vitro diagnostic use in the quantitative measurement of microalbumin in human urine on ADVIA Chemistry systems/ Atellica CH and Atellica CI Analyzers; UDI-DI: 00630414596310 Lot Numbers: 232033, 232128, 232137, 232146, 232147, 242149, 242150, 24236518483 units
Z-3170-2024ADVIA Chemistry Microalbumin_2 (¿ALB_2) -For in vitro diagnostic use in the quantitative measurement of microalbumin in human urine on ADVIA Chemistry systems/ Atellica CH and Atellica CI Analyzers.; UDI-DI: 00630414517643 Lot Numbers: 612528, 616644, 620971, 623768, 636498, 640513, 658356, 66887219645 units
Based on the official FDA remedy

What to do — step by step

  1. Check Kit Lot Number

    Look for the lot numbers 232033, 232128, 232137, 232146, 232147, 242149, 242150, or 242365 on the kit packaging.

Where it went and when

Where it was distributed

Distributed nationwide, and outside the U.S.

Alaska: yesAKMaine: yesMEVermont: yesVTNew Hampshire: yesNHWashington: yesWAIdaho: yesIDMontana: yesMTNorth Dakota: yesNDMinnesota: yesMNIllinois: yesILWisconsin: yesWIMichigan: yesMINew York: yesNYRhode Island: yesRIMassachusetts: yesMAOregon: yesORNevada: yesNVWyoming: yesWYSouth Dakota: yesSDIowa: yesIAIndiana: yesINOhio: yesOHPennsylvania: yesPANew Jersey: yesNJConnecticut: yesCTCalifornia: yesCAUtah: yesUTColorado: yesCONebraska: yesNEMissouri: yesMOKentucky: yesKYWest Virginia: yesWVVirginia: yesVAMaryland: yesMDDelaware: yesDEArizona: yesAZNew Mexico: yesNMKansas: yesKSArkansas: yesARTennessee: yesTNNorth Carolina: yesNCSouth Carolina: yesSCDistrict of Columbia: yesDCOklahoma: yesOKLouisiana: yesLAMississippi: yesMSAlabama: yesALGeorgia: yesGAHawaii: yesHITexas: yesTXFlorida: yesFL

FDA distribution: “Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belgium, Bosnia Herzeg, Brazil, Canada, Chile, Colombia, Croatia, Cura¿ao, St. Eus, Czech Republic, Denmark, Ecuador,…” States read from FDA’s own wording; “not listed” means the notice does not name it.

Distributed nationwide, and outside the U.S.

Recall timeline

FDA published this recall 41 days after the firm started it.

1. Firm started the recall: Aug 15, 202412. FDA classified it Class II: Sep 18, 202423. Published in FDA’s Enforcement Report: Sep 25, 202432024
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-3169-2024 (August 15, 2024) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-3169-2024 to see it in context.
  • Plain-language summary written by qwen3:4b on August 19, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA