Use caution Medical device recall · FDA · Published August 15, 2024

B Braun SL Bloodline Tubing Recall: Risk of Occlusion During Dialysis

B Braun recalls SL Bloodline tubing for FMC 2008 series to prevent occlusion during hemodialysis. Incomplete connector insertion may cause flow restrictions, requiring switch to retrograde flow.

🇺🇸 FDA recall #Z-3173-2024Medical deviceB Braun ›
Take action

Contact the company

Recalling firmB Braun Medical Inc824 12th Ave · Bethlehem, PA 18018-3524
✎ GovInfoNow plain-language explainer

B Braun is recalling SL Bloodline tubing for FMC 2:008 series used in hemodialysis. If the patient connector isn't fully inserted into the Locksite, it may cause flow restrictions and delays in treatment. Read more at govinfonow.com/recall/54217-b-braun-sl-bloodline-tubing-recall-risk-of-occlusion-during.

Source Summary by GovInfoNow · govinfonow.com/recall/54217-b-braun-sl-bloodline-tubing-recall-risk-of-occlusion-during · Updated August 19, 2026 · FDA enforcement record: FDA #Z-3173-2024

Cite this page

“B Braun SL Bloodline Tubing Recall: Risk of Occlusion During Dialysis.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/54217-b-braun-sl-bloodline-tubing-recall-risk-of-occlusion-during. Based on FDA recall #Z-3173-2024 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-3173-2024%22&limit=1).

The FDA announced on August 15, 2024, that B Braun Medical Inc is recalling the SL Bloodline for FMC 2008 Series due to incomplete patient connector insertion. This defect may cause occlusion during prograde rinseback procedures and delay therapy. The agency classified the recall as Class II.

How to tell if you have it

  • Brand: B Braun
  • Product: SL Bloodline for FMC 2008 Series-Tubing
  • Model/Catalog Number: SL-2000M2095
  • UDI-DI: 04046964367762
  • Lot Codes: All Unexpired Lot Numbers
  • Units Affected: 10,658,351

The hazard

Incomplete insertion of the patient connector of the Streamline Airless System Hemodialysis Bloodline to the Locksite may result in the potential for occlusion during prograde rinseback procedures. If the patient connector is not fully inserted into the Locksite, the internal septum may not open fully enough to prevent restrictions to flow. This may cause delays in therapy while clinicians are forced to switch to retrograde flow method.

Context

GovInfoNow's records show B Braun Medical Inc has issued 66 other recalls since October 15, 2013. Recent recalls include a sodium chloride injection recall on August 19, 2026, and a lactated ringer's injection recall on August 12, 2026, both citing particulate matter hazards. Other recent recalls of similar products include Fresenius Medical Care 5008X Blood Tubing Set Recall (August 10, 2026) and Fresenius 5008X HD/HDF with CLiC Blood Tubing Set Recall (August 10, 2026).

How to tell if you have this one

Check for model number SL-2000M2095 on the tubing. The product was sold with UDI-DI 04046964367762 and all unexpired lot numbers.

  • Brand
    B Braun
  • Category
    Medical device
FDA recall event · 7 products

All 7 products in this recall

FDA lists each product line separately under one recall event. Check your product description, lot or code numbers against each row. Every row shares the hazard and remedy on this page.

  • Classification
    Class II
  • Status
    Ongoing
  • Distribution
    US Nationwide distribution.
Recall numberProduct, codes & lotsQuantity
Z-3173-2024SL BLOODLINE FOR FMC 2008 SERIES-Tubing to be used to transfer blood between a patient and a haemodialyser during haemodialysis. Model/Catalog Number: SL-2000M2095; UDI-DI 04046964367762; All Unexpired Lot Numbers10658351 units
Z-3174-2024STREAMLINE FRESENIUS FOR DAVITA-Tubing to be used to transfer blood between a patient and a haemodialyser during haemodialysis. Model/Catalog Number: SL-2000M2095D; UDI-DI 04046955348763; All Unexpired Lot Numbers4671540 units
Z-3175-2024STREAMLINE FRESENIUS FOR DAVITA-Tubing to be used to transfer blood between a patient and a haemodialyser during haemodialysis. Model/Catalog Number: SL-2000M2095DA; UDI-DI 04046955776023; All Unexpired Lot Numbers4984920 units
Z-3176-2024STREAMLINE BLOODLINE LONG VERSION, FMC-Tubing to be used to transfer blood between a patient and a haemodialyser during haemodialysis. Model/Catalog Number: SL-2000M2095L; UDI-DI 04046955348909; All Unexpired Lot Numbers2021376 units
Z-3177-2024STREAMLINE LONG-Tubing to be used to transfer blood between a patient and a haemodialyser during haemodialysis. Model/Catalog Number: SL-2000M2095W; UDI-DI 04046955348800; All Unexpired Lot Numbers142336 units
Z-3178-2024STREAMLINE BLOODLINE SET FOR DIALOG-Tubing to be used to transfer blood between a patient and a haemodialyser during haemodialysis. Model/Catalog Number: SL-2010M2096; UDI-DI 04046964367786; All Unexpired Lot Numbers9864165 units
Z-3179-2024STREAMLINE BLOODLINE SET FOR DIALOG DR-Tubing to be used to transfer blood between a patient and a haemodialyser during haemodialysis. Model/Catalog Number: SL-2010M2096A; UDI-DI 04046955674992; All Unexpired Lot Numbers11262112 units
Based on the official FDA remedy

What to do — step by step

  1. Check Connector Insertion

    Ensure the patient connector is fully inserted into the Locksite before use to prevent flow restrictions.

Where it went and when

Where it was distributed

Distributed nationwide.

Alaska: yesAKMaine: yesMEVermont: yesVTNew Hampshire: yesNHWashington: yesWAIdaho: yesIDMontana: yesMTNorth Dakota: yesNDMinnesota: yesMNIllinois: yesILWisconsin: yesWIMichigan: yesMINew York: yesNYRhode Island: yesRIMassachusetts: yesMAOregon: yesORNevada: yesNVWyoming: yesWYSouth Dakota: yesSDIowa: yesIAIndiana: yesINOhio: yesOHPennsylvania: yesPANew Jersey: yesNJConnecticut: yesCTCalifornia: yesCAUtah: yesUTColorado: yesCONebraska: yesNEMissouri: yesMOKentucky: yesKYWest Virginia: yesWVVirginia: yesVAMaryland: yesMDDelaware: yesDEArizona: yesAZNew Mexico: yesNMKansas: yesKSArkansas: yesARTennessee: yesTNNorth Carolina: yesNCSouth Carolina: yesSCDistrict of Columbia: yesDCOklahoma: yesOKLouisiana: yesLAMississippi: yesMSAlabama: yesALGeorgia: yesGAHawaii: yesHITexas: yesTXFlorida: yesFL

FDA distribution: “US Nationwide distribution.” States read from FDA’s own wording; “not listed” means the notice does not name it.

Distributed nationwide.

Recall timeline

FDA published this recall 41 days after the firm started it.

1. Firm started the recall: Aug 15, 202412. FDA classified it Class II: Sep 19, 202423. Published in FDA’s Enforcement Report: Sep 25, 202432024
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-3173-2024 (August 15, 2024) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-3173-2024 to see it in context.
  • Plain-language summary written by qwen3:4b on August 19, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA